The effect of repeated torque in small diameter implants with machined and premachined abutments.

Abolfazl Saboury1, Hamid Neshandar Asli, Shahram Vaziri.   

Abstract

BACKGROUND: Detorquing value is an important factor in the amount of preload stresses during abutment screw fastening. This study evaluated the percentage of detorque values in two-piece machined titanium and premachined cast abutments in small diameter implants.
MATERIALS AND METHODS: Three groups of five samples were evaluated. Group 1 (G1), machined titanium abutments, group 2 (G2), premachined cast straight abutments that cast with gold-palladium, and group 3 (G3), premachined angled cast abutments that cast with the same alloy, were angled before casting. Each abutment was torque to 24 Ncm according to the manufacturer's instructions and detorqued five times. The means of detorquing and torquing values in all groups were recorded. The mean of detorque in each group as a percentage of the toque value was calculated. The data for all groups were compared and calculated using analysis of variance (ANOVA) and t-test.
RESULTS: Mean detorque values in G1, G2, and G3 were 88.1 ± 1.69, 93.1 ± 2.68, and 80.9 ± 4.95%, respectively. The ANOVA showed significant differences in mean of applied detorque (p < .001) and torque (p = .06) tightening among different groups. G2 had significantly greater detorque values (p < .05). No significant differences were found between G1 and G2. Surprisingly, abutment screw fracture occurred in three samples of G3.
CONCLUSIONS: G3 showed significant percentage torque reduction (p < .05) and exhibited abutment screw fracture during evaluation. G2 presented the lowest torque reduction. Screw fracture occurred only in G3.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22376004     DOI: 10.1111/j.1708-8208.2012.00442.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  3 in total

1.  In Vitro Study of Preload Loss in Different Implant Abutment Connection Designs.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Marta Relvas; Ana Cristina Braga; Blanca Ríos-Carrasco; Javier Gil; José Vicente Rios-Santos; Ana Fernández-Palacín; Mariano Herrero-Climent
Journal:  Materials (Basel)       Date:  2022-02-14       Impact factor: 3.623

2.  The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections.

Authors:  Fariba Saleh Saber; Nader Abolfazli; Soheil Jannatii Ataei; Mahsa Taghizade Motlagh; Vahede Gharekhani
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-06-21

Review 3.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.