| Literature DB >> 22339804 |
Yulin Qi1, Mark P Barrow, Huilin Li, Joseph E Meier, Steve L Van Orden, Christopher J Thompson, Peter B O'Connor.
Abstract
The Fourier transform spectrum can be presented in the absorption-mode (commonly used in FT-NMR), magnitude-mode (FT-ICR), and power-mode (engineering applications). As is routinely used in FT-NMR, it is well-known that the absorption-mode display gives a much narrower peak shape which greatly improves the spectrum; recently, the successful solution of the phase equation allowed broadband phase correction which makes it possible to apply the absorption-mode routinely in FT-ICR. With the empirical evidence provided herein, it has been confirmed that in addition to the improvement on resolving power, compared to the conventional magnitude-mode, the new absorption-mode improves the signal-to-noise ratio (S/N) of a spectrum by 1.4-fold and can improve the mass accuracy up to 2-fold with no extra cost in instrumentation. Therefore, it is worthwhile to apply and promote absorption-mode in routine FT-ICR experiments.Year: 2012 PMID: 22339804 DOI: 10.1021/ac3000122
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986