Evaluation of a digital flocculation reader for the rapid plasma reagin test for the serological diagnosis of syphilis.

Arnold R Castro1, David D Binks, Danny L Raymer, Susan E Kikkert, Heather A Jost, Mahin M Park, Benjamin D Card, David L Cox.   

Abstract

We described the ASiManager-AT digital flocculation reader to demonstrate concordance between visual and digital readings of the rapid plasma reagin test for detection of antibodies in the serum of patients with syphilis. A qualitative and quantitative rapid plasma reagin was performed on each serum samples giving a concordance of 98.6% and 99.7%, respectively, for reactives and 100% for nonreactives.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22337110     DOI: 10.1097/OLQ.0b013e3182389ab9