Design of a wideband multi-channel system for time reversal hyperthermia.

Hana Dobšíček Trefná1, Paolo Togni, Reza Shiee, Jan Vrba, Mikael Persson.   

Abstract

PURPOSE: To design and test a wideband multi-channel amplifier system for time reversal (TR) microwave hyperthermia, operating in the frequency range 300 MHz-1 GHz, enabling operation in both pulsed and continuous wave regimes. This is to experimentally verify that adaptation of the heating pattern with respect to tumour size can be realised by varying the operating frequency of the antennas and potentially by using Ultra-wideband (UWB) pulse sequences instead of pure harmonic signals.
MATERIALS AND METHODS: The proposed system consists of 12 identical channels driven by a common reference signal. The power and phase settings are applied with resolutions of 0.1 W and 0.1°, respectively. Using a calibration procedure, the measured output characteristics of each channel are interpolated using polynomial functions, which are then implemented into a system software algorithm driving the system feedback loop.
RESULTS: The maximum output power capability of the system varies with frequency, between 90 and 135 W with a relative power error of ± 6%. A phase error in the order of ± 4° has been achieved within the entire frequency band.
CONCLUSIONS: The developed amplifier system prototype is capable of accurate power and phase delivery, over the entire frequency band of the system. The output power of the present system allows for an experimental verification of a recently developed TR-method on phantoms or animals. The system is suitable for further development for head and neck tumours, breast or extremity applications.

Entities:  

Mesh:

Year:  2012        PMID: 22335231     DOI: 10.3109/02656736.2011.641655

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  4 in total

1.  Benefit of replacing the Sigma-60 by the Sigma-Eye applicator. A Monte Carlo-based uncertainty analysis.

Authors:  R A M Canters; M M Paulides; M Franckena; J W Mens; G C van Rhoon
Journal:  Strahlenther Onkol       Date:  2012-11-18       Impact factor: 3.621

Review 2.  Simulation techniques in hyperthermia treatment planning.

Authors:  Margarethus M Paulides; Paul R Stauffer; Esra Neufeld; Paolo F Maccarini; Adamos Kyriakou; Richard A M Canters; Chris J Diederich; Jurriaan F Bakker; Gerard C Van Rhoon
Journal:  Int J Hyperthermia       Date:  2013-05-14       Impact factor: 3.914

3.  Hyperthermia Treatment Planning Including Convective Flow in Cerebrospinal Fluid for Brain Tumour Hyperthermia Treatment Using a Novel Dedicated Paediatric Brain Applicator.

Authors:  Gerben Schooneveldt; Hana Dobšíček Trefná; Mikael Persson; Theo M de Reijke; Klas Blomgren; H Petra Kok; Hans Crezee
Journal:  Cancers (Basel)       Date:  2019-08-15       Impact factor: 6.639

4.  Design, Implementation, Evaluation and Application of a 32-Channel Radio Frequency Signal Generator for Thermal Magnetic Resonance Based Anti-Cancer Treatment.

Authors:  Haopeng Han; Thomas Wilhelm Eigentler; Shuailin Wang; Egor Kretov; Lukas Winter; Werner Hoffmann; Eckhard Grass; Thoralf Niendorf
Journal:  Cancers (Basel)       Date:  2020-06-28       Impact factor: 6.639

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.