The endothelial sample size analysis in corneal specular microscopy clinical examinations.

Fernando C Abib1, Ricardo Holzchuh, Artur Schaefer, Tania Schaefer, Ronialci Godois.   

Abstract

PURPOSE: To evaluate endothelial cell sample size and statistical error in corneal specular microscopy (CSM) examinations.
METHODS: One hundred twenty examinations were conducted with 4 types of corneal specular microscopes: 30 with each BioOptics, CSO, Konan, and Topcon corneal specular microscopes. All endothelial image data were analyzed by respective instrument software and also by the Cells Analyzer software with a method developed in our lab. A reliability degree (RD) of 95% and a relative error (RE) of 0.05 were used as cut-off values to analyze images of the counted endothelial cells called samples. The sample size mean was the number of cells evaluated on the images obtained with each device. Only examinations with RE < 0.05 were considered statistically correct and suitable for comparisons with future examinations. The Cells Analyzer software was used to calculate the RE and customized sample size for all examinations.
RESULTS: Bio-Optics: sample size, 97 ± 22 cells; RE, 6.52 ± 0.86; only 10% of the examinations had sufficient endothelial cell quantity (RE < 0.05); customized sample size, 162 ± 34 cells. CSO: sample size, 110 ± 20 cells; RE, 5.98 ± 0.98; only 16.6% of the examinations had sufficient endothelial cell quantity (RE < 0.05); customized sample size, 157 ± 45 cells. Konan: sample size, 80 ± 27 cells; RE, 10.6 ± 3.67; none of the examinations had sufficient endothelial cell quantity (RE > 0.05); customized sample size, 336 ± 131 cells. Topcon: sample size, 87 ± 17 cells; RE, 10.1 ± 2.52; none of the examinations had sufficient endothelial cell quantity (RE > 0.05); customized sample size, 382 ± 159 cells.
CONCLUSIONS: A very high number of CSM examinations had sample errors based on Cells Analyzer software. The endothelial sample size (examinations) needs to include more cells to be reliable and reproducible. The Cells Analyzer tutorial routine will be useful for CSM examination reliability and reproducibility.

Entities:  

Mesh:

Year:  2012        PMID: 22333658     DOI: 10.1097/ICO.0b013e3181cc7961

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  9 in total

1.  Corneal endothelial cell analysis using two non-contact specular microscopes in healthy subjects.

Authors:  Manuel Garza-Leon
Journal:  Int Ophthalmol       Date:  2015-10-05       Impact factor: 2.031

2.  Outcomes of corneal collagen crosslinking using a customized epithelial debridement technique in keratoconic eyes with thin corneas.

Authors:  Nurullah Cagil; Ozge Sarac; Gamze Dereli Can; Emine Akcay; Mehmet Erol Can
Journal:  Int Ophthalmol       Date:  2016-04-21       Impact factor: 2.031

3.  Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes.

Authors:  Irmak Karaca; Suzan Guven Yilmaz; Melis Palamar; Halil Ates
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

4.  Age-related changes of corneal endothelial cell in healthy Chinese tree shrew measured by non-contact specular microscope.

Authors:  Min Wu; De-Xuan Kuang; Ya-Qi Huang; Yu-Run Miao; Xiao-Cheng Liu; Jie-Jie Dai
Journal:  Int J Ophthalmol       Date:  2017-12-18       Impact factor: 1.779

5.  Automated segmentation of the corneal endothelium in a large set of 'real-world' specular microscopy images using the U-Net architecture.

Authors:  Moritz C Daniel; Lisa Atzrodt; Felicitas Bucher; Katrin Wacker; Stefan Böhringer; Thomas Reinhard; Daniel Böhringer
Journal:  Sci Rep       Date:  2019-03-18       Impact factor: 4.379

6.  Improved Interchangeability with Different Corneal Specular Microscopes for Quantitative Endothelial Cell Analysis.

Authors:  Gwyneth A van Rijn; C Jasper F Wijnen; Bart Th van Dooren; Yanny Yy Cheng; Jan-Willem M Beenakker; Gregorius Pm Luyten
Journal:  Clin Ophthalmol       Date:  2020-01-13

7.  Cell-by-cell alignment of repeated specular microscopy images from the same eye.

Authors:  Daniel Böhringer; Stefan Lang; Thomas Reinhard
Journal:  PLoS One       Date:  2013-03-14       Impact factor: 3.240

8.  Regulation of functional corneal endothelial cells isolated from sphere colonies by Rho-associated protein kinase inhibitor.

Authors:  Yan-Long Bi; Qi Zhou; Fei DU; Ming-Feng Wu; Guo-Tong Xu; Gui-Qin Sui
Journal:  Exp Ther Med       Date:  2012-11-19       Impact factor: 2.447

9.  Comparison of corneal endothelial cell measurements by two non-contact specular microscopes.

Authors:  Laura Gasser; Thomas Reinhard; Daniel Böhringer
Journal:  BMC Ophthalmol       Date:  2015-07-29       Impact factor: 2.209

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.