| Literature DB >> 22333069 |
Ling Wang1, Michael Riss, Jennifer Olmos Buitrago, Enric Claverol-Tinturé.
Abstract
We have previously described the use of microchannels (μChannels) as substrate-integrated equivalents of micropipettes and advantageous neuron-electrode interface enhancers. The use of μChannels to establish stable recording and stimulation of threading axons results in a high signal-to-noise ratio (SNR), potentially high-throughput and low-cost alternative to conventional substrate-embedded microelectrodes. Here we confirm the consistent achievement of high SNRs with μChannels and systematically characterize the impact of μChannel geometry on the measured signals via numerical simulations and in vitro experiments. We demonstrate and rationalize how channels with a length of ≤300 μm and channel cross section of ≤12 μm(2) support spontaneous formation of seals and yield spike sizes in the millivolt range. Despite the low degree of complexity involved in their fabrication and use, μChannel devices provide a single-unit mean SNR of 101 ± 76, which compares favourably with the SNR obtained from typical microelectrode arrays.Mesh:
Year: 2012 PMID: 22333069 DOI: 10.1088/1741-2560/9/2/026010
Source DB: PubMed Journal: J Neural Eng ISSN: 1741-2552 Impact factor: 5.379