High-resolution ultrasound in the detection of silicone gel breast implant shell failure: background, in vitro studies, and early clinical results.

Bradley P Bengtson1, Felmont F Eaves.   

Abstract

BACKGROUND: Magnetic resonance imaging (MRI) has historically been considered the "gold standard" for imaging silicone gel breast implants and is currently recommended by the US Food and Drug Administration for device surveillance. Recent studies, however, have questioned its accuracy as the best screening test for implant failure. In addition, the high cost of MRI is a significant deterrent to follow-up, especially among asymptomatic patients. Recent advancements in ultrasound technology have led to the development of high-resolution devices with the potential to accurately image breast implants and breast tissues.
OBJECTIVES: The authors evaluate the feasibility of portable, high-resolution ultrasound (HRUS) for imaging of silicone gel breast implants and perform preliminary comparisons of HRUS to MRI in the assessment of both intact and failed implants in a clinical setting by both radiologists and plastic surgeons.
METHODS: Phase 1 was composed of in vitro and ex vivo scanning model assessments in a variety of implant models utilizing multiple HRUS hardware platforms (GE LOGIQ-9, LOGIQ-e, LOGIQ-i, and Venue-40 devices) and transducer heads (range, 8-16 MHz, mainly GE12ML transducer). In Phase 2, these technologies were applied clinically to provide imaging experience in three patients previously diagnosed with unilateral implant failure. Phase 3 was a preliminary prospective evaluation of HRUS of 29 implants in 15 consecutive patients for whom MRI and independent surgeon-performed and radiologist-performed HRUS scans were compared to subsequent surgical findings.
RESULTS: In Phase 1, all hardware models easily detected both intact and intentionally damaged shells in currently marketed fourth-generation responsive gel implants and in investigational, fifth-generation highly-cohesive gel devices. Although multiple transducers were able to detect shell failure, the 12-MHz head produced the best images at the normal clinical depth range. In Phase 2, confirmatory HRUS scans correctly identified the side of rupture and were consistent with MRI and surgical findings in all patients. In Phase 3, MRI, surgeon-performed HRUS, and radiologist-performed HRUS scans were all accurate in predicting implant shell integrity in 29 of 29 imaged breasts (100%) as confirmed at the time of surgery in both symptomatic and asymptomatic patients.
CONCLUSIONS: Preliminary results with a variety of base and transducer systems demonstrated that HRUS provides excellent visualization of current fourth- and fifth-generation silicone gel implants both in the in vitro and ex vivo scanning models. In vivo surgeon-performed HRUS accurately identified implant status and correlated with radiologist-performed HRUS, MRI, and surgical findings. An ongoing Phase 4 prospective study is under way to help define the sensitivity and specificity of HRUS technologies in the evaluation of current implant designs. However, the relative affordability, accessibility, availability, and dynamic real-time visualization provided by HRUS represent significant potential advantages of HRUS over MRI in both the screening and future diagnosis of breast implant shell failure.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22328687     DOI: 10.1177/1090820X11434507

Source DB:  PubMed          Journal:  Aesthet Surg J        ISSN: 1090-820X            Impact factor:   4.283


  7 in total

Review 1.  How to prevent complications in breast augmentation.

Authors:  Maurizio Bruno Nava; Alberto Rancati; Claudio Angrigiani; Giuseppe Catanuto; Nicola Rocco
Journal:  Gland Surg       Date:  2017-04

2.  A Surgeon's Empirical Perspectives on Use of High-resolution Ultrasound in Preoperatively Detecting a Rupture in the Context of Breast Implant Crisis in Korea.

Authors:  Bum Sik Bang; Seong Hoon Jung; Eun Kyoung Lee; Jung Youp Sung; Keun Yeong Song; Young Bum Yoo; Dong Wook Park; Jeong Eun Sohn; Jae Hong Kim
Journal:  Aesthetic Plast Surg       Date:  2022-03-16       Impact factor: 2.708

3.  The Role of High Resolution Ultrasonography in Elucidating Features of the Breast Implants in Asymptomatic Patients After Implant-based Augmentation Mammaplasty.

Authors:  Sang Eun Nam; Bum Sik Bang; Eun Kyoung Lee; Jung Youp Sung; Keun Yeong Song; Young Bum Yoo; Dong Wook Park; Jae Hong Kim
Journal:  Aesthetic Plast Surg       Date:  2022-01-13       Impact factor: 2.708

4.  Safety and Effectiveness of Silicone Gel-Filled Breast Implants in Primary Augmentation Patients.

Authors:  Daniel Maxwell; Megan Estes; Jennifer McMillen Walcott; John W Canady; Tina D Hunter; Larry Gache; Bernadette Wang-Ashraf; Diane Alexander
Journal:  Aesthet Surg J       Date:  2021-05-18       Impact factor: 4.283

5.  Silent Rupture of Silicone Gel Breast Implants: High-Resolution Ultrasound Scans and Surveys of 584 Women.

Authors:  Marc J Salzman
Journal:  Plast Reconstr Surg       Date:  2022-01-01       Impact factor: 4.730

6.  The Expanding Role of Diagnostic Ultrasound in Plastic Surgery.

Authors:  Eric Swanson
Journal:  Plast Reconstr Surg Glob Open       Date:  2018-09-05

7.  A Preliminary Retrospective Study to Assess the Short-Term Safety of Traditional Smooth or Microtextured Silicone Gel-Filled Breast Implants in Korea.

Authors:  Sanghyuk Han; Robert Kim; Tae Seob Kim; Jung Heum Park; Seung Soo Kim; Cheol Jeong; Ji Heui Lee
Journal:  Medicina (Kaunas)       Date:  2021-12-16       Impact factor: 2.430

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.