Literature DB >> 22327674

Telemetric ICP measurement with the first CE-approved device: data from animal experiments and initial clinical experiences.

Michael Kiefer1, Sebastian Antes, Steffen Leonhardt, Melanie Schmitt, Berk Orakcioglu, Oliver W Sakowitz, Regina Eymann.   

Abstract

The objective was to evaluate the qualification of the new telemetric intracranial pressure (ICP) measurement (t-ICP) device Raumedic(®) NEUROVENT P-Tel and S-Tel. The proof of concept was examined in a pilot animal study measuring intraperitoneal pressure with a telemetric and a conventional ICP measurement probe at five rates for 1 h each. Moderate external pressure load allowed measuring values between 0 and 40 mmHg. To estimate long-term performance 18 t-ICP devices were implanted subdurally or intraparenchymally into minipigs. Reference measurements were performed regularly using conventional ICP probes. From the short-term as well as from the long-term perspective t-ICP proved to have excellent dynamic ICP signal components perception (e.g. pulse amplitude). Some zero drift of static ICP was found, ranging between 5 and 8 mmHg. While all telemetric, intraparenchymal probes kept their functionality throughout the follow-up, 33% of the subdurals failed for reasons detailed in another paper. Raumedic's NEUROVENT(®) P-Tel/S-Tel proved to provide reliable data over periods of up to 18 months. Minor zero drift can be well tolerated as the dynamic ICP signal is measured with excellent stability. Clinicians should focus more on such ICP dynamic signal information than on static ICP when using the device over longer follow-up periods.

Entities:  

Mesh:

Year:  2012        PMID: 22327674     DOI: 10.1007/978-3-7091-0956-4_20

Source DB:  PubMed          Journal:  Acta Neurochir Suppl        ISSN: 0065-1419


  9 in total

1.  An operative technique combining endoscopic third ventriculostomy and long-term ICP monitoring.

Authors:  Sebastian Antes; Christoph A Tschan; Joachim M Oertel
Journal:  Childs Nerv Syst       Date:  2013-08-29       Impact factor: 1.475

2.  Feasibility of telemetric ICP-guided valve adjustments for complex shunt therapy.

Authors:  Florian Baptist Freimann; Matthias Schulz; Hannes Haberl; Ulrich-Wilhelm Thomale
Journal:  Childs Nerv Syst       Date:  2013-11-22       Impact factor: 1.475

Review 3.  Microelectromechanical systems and nephrology: the next frontier in renal replacement technology.

Authors:  Steven Kim; Shuvo Roy
Journal:  Adv Chronic Kidney Dis       Date:  2013-11       Impact factor: 3.620

4.  Clinical practice audit concerning antimicrobial prophylaxis in paediatric neurosurgery: results from a German paediatric oncology unit.

Authors:  Katja Weiss; Arne Simon; Norbert Graf; Jakob Schöpe; Joachim Oertel; Stefan Linsler
Journal:  Childs Nerv Syst       Date:  2016-11-07       Impact factor: 1.475

5.  Telemetric intracranial pressure monitoring: a noninvasive method to follow up children with complex craniosynostoses. A case report.

Authors:  Sara Magnéli; Timothy Howells; Daniel Saiepour; Daniel Nowinski; Per Enblad; Pelle Nilsson
Journal:  Childs Nerv Syst       Date:  2016-02-10       Impact factor: 1.475

Review 6.  Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.

Authors:  Usmah Kawoos; Richard M McCarron; Charles R Auker; Mikulas Chavko
Journal:  Int J Mol Sci       Date:  2015-12-04       Impact factor: 5.923

7.  Single center experiences with telemetric intracranial pressure measurements in patients with CSF circulation disturbances.

Authors:  Valentina Pennacchietti; Vincent Prinz; Andreas Schaumann; Tobias Finger; Matthias Schulz; Ulrich W Thomale
Journal:  Acta Neurochir (Wien)       Date:  2020-06-03       Impact factor: 2.216

8.  Benefits and obstacles of telemetric ICP monitoring.

Authors:  Joachim M K Oertel; Matthias J M Huelser
Journal:  Acta Neurochir (Wien)       Date:  2021-02-07       Impact factor: 2.216

Review 9.  Measuring intracranial pressure by invasive, less invasive or non-invasive means: limitations and avenues for improvement.

Authors:  Karen Brastad Evensen; Per Kristian Eide
Journal:  Fluids Barriers CNS       Date:  2020-05-06
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.