Intracranial pressure telemetry: first experience of an experimental in vivo study using a new device.

Berk Orakcioglu1, Christopher Beynon, Modar M Kentar, Regina Eymann, Michael Kiefer, Oliver W Sakowitz.   

Abstract

OBJECTIVE: To test two new telemetric intracranial pressure (ICP) probes (NEUROVENT(®)-P-tel, NEUROVENT(®)-S-tel) in a porcine model. We aimed to intraoperatively correlate the telemetric probes to parenchymal ICP probes and study their reliability in the first hours after implantation. The experimental set-up, new telemetric technology and first data will be presented.
METHODS: We implanted a right parietal (parenchymal) and left parietal (subdural) telemetric ICP probe in 13 Göttingen mini-pigs under general anaesthesia. Through the left parietal burr hole a parenchymal ICP probe (Neurovent(®) ICP) was introduced. Intraoperatively, the head position was changed to provoke ICP changes every 10 min. The telemetric probes were left in situ and finally the parenchymal ICP probe was removed. We correlated mean differences between each telemetric probe and the conventional ICP measurement and Bland-Altman plots were generated for statistical analysis.
RESULTS: We present first data containing intraoperative measurements of 26 telemetric probes after implantation. Intraoperatively, mean differences of 2.48 ± 1.52 mmHg SD (NEUROVENT(®)-P-tel) and 2.64 ± 1.79 mmHg (NEUROVENT(®)-S-tel) were observed. The Bland-Altman plot demonstrates good correlation of the telemetric probes compared with parenchymal ICP probes.
CONCLUSION: We present a new telemetric technology that was experimentally compared with a parenchymal ICP probe. We provide data that the new telemetric probes will comparably measure ICP vs an external ICP probe. This stand-alone ICP tool may allow permanent measurement of ICP in hydrocephalus patients. Further continuation of our study will demonstrate whether this system guarantees acceptable long-term reliability.

Entities:  

Mesh:

Year:  2012        PMID: 22327673     DOI: 10.1007/978-3-7091-0956-4_19

Source DB:  PubMed          Journal:  Acta Neurochir Suppl        ISSN: 0065-1419


  4 in total

1.  An operative technique combining endoscopic third ventriculostomy and long-term ICP monitoring.

Authors:  Sebastian Antes; Christoph A Tschan; Joachim M Oertel
Journal:  Childs Nerv Syst       Date:  2013-08-29       Impact factor: 1.475

2.  Feasibility of telemetric ICP-guided valve adjustments for complex shunt therapy.

Authors:  Florian Baptist Freimann; Matthias Schulz; Hannes Haberl; Ulrich-Wilhelm Thomale
Journal:  Childs Nerv Syst       Date:  2013-11-22       Impact factor: 1.475

Review 3.  Microelectromechanical systems and nephrology: the next frontier in renal replacement technology.

Authors:  Steven Kim; Shuvo Roy
Journal:  Adv Chronic Kidney Dis       Date:  2013-11       Impact factor: 3.620

4.  [Non-invasive estimation of intracranial pressure : MR-based evaluation in children with hydrocephalus].

Authors:  M Muehlmann; D Steffinger; A Peraud; M Lehner; F Heinen; N Alperin; B Ertl-Wagner; I K Koerte
Journal:  Radiologe       Date:  2012-09       Impact factor: 0.635

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.