Evaluation of the impedance analyzer PocH-100iV Diff for analysis of canine and feline blood.

Natali B Bauer1, Julia Nakagawa, Cathrin Dunker, Klaus Failing, Andreas Moritz.   

Abstract

BACKGROUND: An automated impedance-based in-house hematology analyzer, the PocH-100iV Diff, which provides a 3-part leukocyte differential count that includes eosinophils, recently has been introduced.
OBJECTIVES: The aims of this study were to validate results from the PocH-100iV Diff for dogs and cats and evaluate the impact of the anticoagulant used and sample storage conditions.
METHODS: Blood samples collected in K(3) EDTA from 153 cats and 150 dogs were included in the comparison study. The reference analyzer was the ADVIA 2120 hematology analyzer, and manual differential leukocyte counts and PCV were the manual reference methods.
RESULTS: Coefficients of variation were < 3% except for platelet counts and feline differential and eosinophil counts. Correlation between analyzers was good to excellent except for hemoglobin (HGB) concentration in dogs and RBC indices for both species. Biases were close to 0 except for MCHC and platelet counts. Correlation with manual counts was good for lymphocytes and OTHR cells (combined neutrophil and monocyte counts) and fair and poor for feline and canine eosinophil counts, respectively. Estimated sensitivity and specificity for detection of eosinophilia were, respectively, 50% and 98% for cats and 34% and 77% for dogs. A significant anticoagulant effect was seen for MCV in cats and for HCT, MCH, MCHC, and platelet, OTHR, and eosinophil counts in dogs. RBC and WBC counts, HGB concentration, and MCH were stable for 72 h. HCT, MCV, MCHC, and platelet counts were affected by sample storage (dogs > cats; 22°C > 4°C).
CONCLUSIONS: The PocH-100iV Diff is a suitable in-house instrument. A strength is its specific, but moderately sensitive, detection of feline eosinophils.
© 2012 American Society for Veterinary Clinical Pathology.

Entities:  

Mesh:

Year:  2012        PMID: 22324888     DOI: 10.1111/j.1939-165X.2012.00405.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  4 in total

1.  Rapid, effective and user-friendly immunophenotyping of canine lymphoma using a personal flow cytometer.

Authors:  Stratos Papakonstantinou; Inese Berzina; Amanda Lawlor; Emma J O'Neill; Peter J O'Brien
Journal:  Ir Vet J       Date:  2013-04-01       Impact factor: 2.146

2.  Hematologic and serum biochemical reference intervals of the Oriental white stork (Ciconia boyciana) and the application of an automatic hematologic analyzer.

Authors:  Jae-Ik Han; Hye-Jin Jang; Ki-Jeong Na
Journal:  J Vet Sci       Date:  2016-09-30       Impact factor: 1.672

3.  Development and evaluation of a novel in-clinic automated hematology analyzer, ProCyte Dx, for canine erythrocyte indices, leukogram, platelet counts and reticulocyte counts.

Authors:  Yasuhito Fujino; Yoichi Nakamura; Hideaki Matsumoto; Kenjiro Fukushima; Masashi Takahashi; Koichi Ohno; Hajime Tsujimoto
Journal:  J Vet Med Sci       Date:  2013-07-01       Impact factor: 1.267

4.  Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats.

Authors:  Kim-Lina Charlotte Zelmer; Natali Bauer; Andreas Moritz
Journal:  J Vet Diagn Invest       Date:  2022-03-24       Impact factor: 1.279

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.