Biomechanical testing of suture-based meniscal repair devices containing ultrahigh-molecular-weight polyethylene suture: update 2011.

F Alan Barber1, Morley A Herbert, Eric D Bava, Otis R Drew.   

Abstract

PURPOSE: To evaluate the biomechanical characteristics of recently introduced ultrahigh-molecular-weight polyethylene suture-based, self-adjusting meniscal repair devices.
METHODS: Updating a prior study published in 2009, we made vertical longitudinal cuts 3 mm from the periphery in fresh-frozen adult human menisci to simulate a bucket-handle meniscus tear. Each tear was then repaired by a single repair technique in 10 meniscus specimens. Group 1 menisci were repaired with a vertical mattress suture of No. 2-0 Ethibond (Ethicon, Somerville, NJ). Group 2 menisci were repaired with a vertical mattress suture of No. 2-0 OrthoCord (DePuy Mitek, Raynham, MA). Group 3 menisci were repaired with a single OmniSpan device with No. 2-0 OrthoCord suture (DePuy Mitek). Group 4 menisci were repaired with a single Meniscal Cinch device with No. 2-0 FiberWire suture (Arthrex, Naples, FL). Group 5 menisci were repaired with a single MaxFire device inserted with the MarXmen gun (Biomet Sports Medicine, Warsaw, IN). Group 6 menisci were repaired with a Sequent device with No. 0 Hi-Fi suture (ConMed Linvatec, Largo, FL) in a "V" suture configuration. Group 7 menisci were repaired with a single FasT-Fix 360 device (Smith & Nephew Endoscopy, Andover, MA). By use of a mechanical testing machine, all samples were preloaded at 5 N and cycled 200 times between 5 and 50 N. Those specimens that survived were destructively tested at 5 mm/min. Endpoints included maximum load, displacement, stiffness, and failure mode.
RESULTS: Mean failure loads were as follows: Ethibond suture, 73 N; OrthoCord suture, 88 N; OmniSpan, 88 N; Cinch, 71 N; MarXmen/MaxFire, 54 N; Sequent, 66 N; and FasT-Fix 360, 60 N. Ethibond was stronger than MarXmen/MaxFire. The mean displacement after 100 cycles was as follows: Ethibond, 2.58 mm; OrthoCord, 2.75 mm; OmniSpan, 2.51 mm; Cinch, 2.65 mm; MarXmen/MaxFire, 3.67 mm; Sequent, 3.35 mm; and FasT-Fix 360, 1.13 mm. The MarXmen/MaxFire showed greater 100-cycle displacement than Ethibond and FasT-Fix 360. No difference in stiffness existed for these devices, and failure mode varied without specific trends.
CONCLUSIONS: The biomechanical properties of meniscal repairs using the OmniSpan, Cinch, Sequent, and FasT-Fix 360 devices are equivalent to suture repair techniques. However, the MarXmen/MaxFire meniscal repair device showed significantly lower failure loads and survived less cyclic loading in the human cadaveric meniscus than other tested repairs. CLINICAL RELEVANCE: Most commercially available devices for all-inside meniscal repair using ultrahigh-molecular-weight polyethylene suture provide fixation comparable to the classic vertical mattress suture repair technique in human cadaveric meniscus.
Copyright © 2012 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22317797     DOI: 10.1016/j.arthro.2011.11.020

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  16 in total

1.  All-inside arthroscopic meniscal repair with meniscal cinch.

Authors:  Vipool K Goradia
Journal:  Arthrosc Tech       Date:  2013-05-11

2.  Inside-Out Repair for Radial Meniscus Tears.

Authors:  Ayoosh Pareek; Michael P O'Malley; Bruce A Levy; Michael J Stuart; Aaron J Krych
Journal:  Arthrosc Tech       Date:  2016-07-25

3.  The concept of three-dimensional hold of both circumferential and radial collagen fibres of the meniscus.

Authors:  Ashraf Abdelkafy
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-02-12       Impact factor: 4.342

4.  The effect of inclination angle on the strength of vertical mattress configuration for meniscus repair.

Authors:  Mehmet Erduran; Onur Hapa; Baran Şen; Yavuz Kocabey; Diler Erdemli; Mehran Aksel; Hasan Havitçioğlu
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-04-09       Impact factor: 4.342

5.  Inside-out meniscus repair.

Authors:  Clay G Nelson; Kevin F Bonner
Journal:  Arthrosc Tech       Date:  2013-11-01

Review 6.  Meta-analysis on biomechanical properties of meniscus repairs: are devices better than sutures?

Authors:  Daniel M Buckland; Patrick Sadoghi; Matthias D Wimmer; Patrick Vavken; Geert I Pagenstert; Victor Valderrabano; Claudio Rosso
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-05-20       Impact factor: 4.342

7.  Extra-articular dorsal closing-wedge osteotomy to treat late-stage Freiberg disease using polyblend sutures: technical tips and clinical results.

Authors:  Kazuya Ikoma; Masahiro Maki; Masamitsu Kido; Kan Imai; Yuji Arai; Hiroyoshi Fujiwara; Yasuo Mikami; Toshikazu Kubo
Journal:  Int Orthop       Date:  2014-05-28       Impact factor: 3.075

Review 8.  Meniscus tear surgery and meniscus replacement.

Authors:  Javier Vaquero; Francisco Forriol
Journal:  Muscles Ligaments Tendons J       Date:  2016-05-19

9.  Biomechanical Comparison of Parallel and Crossed Suture Repair for Longitudinal Meniscus Tears.

Authors:  Charles Milchteim; Eric A Branch; Ty Maughon; Jay Hughey; Adam W Anz
Journal:  Orthop J Sports Med       Date:  2016-04-08

10.  Increased Construct Stiffness With Meniscal Repair Sutures and Devices Increases the Risk of Cheese-Wiring During Biomechanical Load-to-Failure Testing.

Authors:  Sebastian Müller; Tanja Schwenk; Michael de Wild; Dimitris Dimitriou; Claudio Rosso
Journal:  Orthop J Sports Med       Date:  2021-06-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.