Literature DB >> 22314805

Eye doses to staff in a nuclear medicine department.

Elizabeth C Summers1, Janis L E Brown, Peter J Bownes, Shona E Anderson.   

Abstract

OBJECTIVE: Occupational radiation doses to the Nuclear Medicine Department staff at Mount Vernon Hospital are routinely measured using optically stimulated luminescence dosemeters for whole-body effective dose and ring thermoluminescence dosemeters (TLDs) for finger dose. In 2002, a project was carried out using LiF:Mg,Cu,P Chinese TLDs to measure the dose to the lens of the eye received by staff during normal working procedures.
METHODS: Separate pairs of TLDs were worn by staff on their forehead between their eyes while dispensing and releasing in the radiopharmacy, injecting, and when administering I-131 capsules to patients. The dose received was calculated using calibration data from identical TLDs irradiated with Tc-99m, I-131, and the Ir-192 source of a Gammamed High Dose Rate (HDR) treatment unit. Data were collected over a 5-month period and the mean dose to the eye was calculated for each procedure.
RESULTS: Using a typical yearly workload, the annual dose to the eye for a single member of staff was calculated and found to be 4.5 mSv.
CONCLUSION: The occupational eye dose limit was, at the time, 150 mSv; therefore, staff were well below the level (3/10th of this limit) that would have required them to be classified. However, there have been large increases in radiopharmacy production and I-131 therapies administered at Mount Vernon in subsequent years. It is therefore expected that the eye dose received by staff will have increased to be significantly higher than 4.5 mSv and will in fact be greater than 6 mSv, which is 3/10th of the proposed new dose limit and would require these staff to become classified workers.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22314805     DOI: 10.1097/MNM.0b013e3283510a8f

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  4 in total

1.  Effective and equivalent dose minimization for personnel in PET procedures: how far are we from the goal?

Authors:  M Lecchi; S Malaspina; A Del Sole
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-12       Impact factor: 9.236

2.  Cataract Risk in a Cohort of U.S. Radiologic Technologists Performing Nuclear Medicine Procedures.

Authors:  Marie-Odile Bernier; Neige Journy; Daphnee Villoing; Michele M Doody; Bruce H Alexander; Martha S Linet; Cari M Kitahara
Journal:  Radiology       Date:  2017-10-11       Impact factor: 11.105

3.  ESTIMATION OF HP(3) AMONG STAFF MEMBERS IN TWO NUCLEAR MEDICINE UNITS IN FINLAND.

Authors:  C Lindholm; A Pekkarinen; O Sipilä; A-L Manninen; M Lehtinen; T Siiskonen
Journal:  Radiat Prot Dosimetry       Date:  2020-08-28       Impact factor: 0.972

4.  Determination of a reliable assessment for occupational eye lens dose in nuclear medicine.

Authors:  Noriaki Miyaji; Kenta Miwa; Takashi Iimori; Kei Wagatsuma; Hiroyuki Tsushima; Noriyo Yokotsuka; Taisuke Murata; Tetsuharu Kasahara; Takashi Terauchi
Journal:  J Appl Clin Med Phys       Date:  2022-07-01       Impact factor: 2.243

  4 in total