Abstract
The present study deals with the estimation by RP-HPLC of two different drug components hydrochlorothiazide and metoprolol tartrate present in a tablet formulation. It is a simple, fast, precise and accurate high performance liquid chromatographic method. It is performed using phosphate buffer along with methanol as mobile phase, in the proportion of 60:40. The separation is done on a C(18) column and it is estimated at a λ(max) of 226 nm with a flow of 1 ml/min. The detection limits range from a 0.013 to 0.075 mg/ml for hydrochlorothiazide and 0.10 to 0.60 mg/ml for metoprolol tartrate, respectively. The specificity for interference of any peak with main peak of interest is checked. A scan of the individual drug was taken for assuring the λ(max). The system suitability by precision is also checked to ensure the analytical method. The method was found to be accurate and precise for estimation of the two drugs simultaneously.Entities:
Keywords: Hydrochlorothiazide; RP-HPLC; metoprolol tartrate
Year: 2011 PMID: 22303067 PMCID: PMC3267308 DOI: 10.4103/0250-474x.91565
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1Typical HPLC chromatogram of hydrochlorothiazide and metoprolol tartrate
The two components were eluted at a retention time of 4.13 min and 10.81 min with an area of 79354 and 136749 for hydrochlorothiazide and metoprolol tartrate respectively. The system suitability was also as per the guidelines and the reproducibility was also checked which was also within the limits.
ASSAY OF COMMERCIAL SAMPLE
PRECISION DATA
RECOVERY DATA
LINEARITY HYDROCHLOROTHIAZIDE
LINEARITY METOPROLOL TARTRATE