| Literature DB >> 22296785 |
Kai Gao1, Chunming Rao, Lei Tao, Chunmei Han, Xinchang Shi, Lan Wang, Wenhong Fan, Lei Yu, Junzhi Wang.
Abstract
This collaborative study characterizes a homogeneous standard for the protein content determination of granulocyte colony-stimulating factor (G-CSF) products with traceability of the measurement. The Kjeldahl method was used to determine the average protein content of G-CSF bulk as 2.505 mg/ml (95% C.I: 2.467-2.543 mg/ml, GCV 4.0%). Using G-CSF bulk as a traceability benchmark, the protein content of the final freeze-dried standard using reverse phase HPLC (RP-HPLC) was 215.4 μg protein per ampoule (95% C.I: 212.407-218.486 μg/ampoule, GCV 3.4%). A comparative study showed that there was no difference between using Filgrastim CRS (European Pharmacopeia G-CSF reference standard) and freeze-dried homogeneous standard when quantifying G-CSF protein content by RP-HPLC (P > 0.05). However, there were significant differences in the G-CSF protein content obtained using a serum albumin standard by Lowry assay and a G-CSF standard with RP-HPLC. Therefore, use of RP-HPLC with a freeze-dried homogeneous standard would eliminate the systematic errors introduced when using a serum albumin standard because of the differences in protein composition between the standard and the sample. It would also be helpful to use this method to compare the quality of G-CSF biosimilar products in situations where the protein content has been calibrated using various standards. Crown Copyright ÂEntities:
Mesh:
Substances:
Year: 2012 PMID: 22296785 DOI: 10.1016/j.biologicals.2011.08.013
Source DB: PubMed Journal: Biologicals ISSN: 1045-1056 Impact factor: 1.856