Literature DB >> 22293438

Cochlear implant electrode array misplacement: a cautionary case report.

E Muzzi1, P Boscolo-Rizzo, R Santarelli, M A Beltrame.   

Abstract

OBJECTIVE: To report a series of pitfalls and complications in a case of cochlear implantation.
METHOD: Case report.
RESULTS: An 11-year-old boy affected by auditory neuropathy underwent cochlear implantation. Intra-operative assessment was apparently consistent with correct insertion of the electrode array into the cochlea. However, subsequent high resolution computed tomography revealed that the entire electrode array was curled up within the vestibule. Revision surgery was complicated by cerebrospinal fluid leakage. A straight probe was repeatedly inserted into the internal auditory canal, before conversion to a canal wall down procedure and appropriate positioning of the electrode array.
CONCLUSION: In this case, mild anteriorisation of the facial nerve created an awkward insertion angle for the electrode array via the retro-facial route, which may have triggered the described series of adverse events.

Entities:  

Mesh:

Year:  2012        PMID: 22293438     DOI: 10.1017/S0022215112000059

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  5 in total

1.  Relationship of cochlea with surrounding neurovascular structures and their implication in cochlear implantation.

Authors:  Daisy Sahni; Anjali Singla; Ashok Gupta; Tulika Gupta; Anjali Aggarwal
Journal:  Surg Radiol Anat       Date:  2015-02-08       Impact factor: 1.246

2.  Extra-Cochlear Insertion in Cochlear Implantation: A Potentially Disastrous Condition.

Authors:  Emine Deniz Gözen; Fırat Tevetoğlu; Haydar Murat Yener; Eyüp Kara; Ahmet Ataş; Sabri Şirolu; Osman Kızılkılıç; Harun Cansız; Emin Karaman
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

3.  The feasibility of endoscopic transcanal approach for insertion of various cochlear electrodes: a pilot study.

Authors:  Lela Migirov; Yisgav Shapira; Michael Wolf
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-03-12       Impact factor: 2.503

4.  Cochlear implant electrode misplacement: incidence, evaluation, and management.

Authors:  Yu-Lan Mary Ying; Jerry W Lin; John S Oghalai; Robert A Williamson
Journal:  Laryngoscope       Date:  2013-01-08       Impact factor: 3.325

5.  Intra-operative radiological diagnosis of a tip roll-over electrode array displacement using fluoroscopy, when electrophysiological testing is normal: the importance of both techniques in cochlear implant surgery.

Authors:  Octavio Garaycochea; Raquel Manrique-Huarte; Manuel Manrique
Journal:  Braz J Otorhinolaryngol       Date:  2017-06-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.