S Serge Barold1, Roland X Stroobandt.
Abstract
This report describes the de novo occurrence of pacemaker-mediated tachycardia (PMT) in a patient with a dual-chamber implantable cardioverter-defibrillator and stable retrograde ventriculoatrial conduction time. The same rate-adaptive post-ventricular atrial refractory period (PVARP) duration had previously prevented PMT. Oversensing of atrial false signals from a defective lead shortened the PVARP with consequent sensing of retrograde conduction.Entities:
Mesh:
Year: 2012 PMID: 22286204 DOI: 10.1093/europace/eur405
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214