Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia.

S Serge Barold1, Roland X Stroobandt.   

Abstract

This report describes the de novo occurrence of pacemaker-mediated tachycardia (PMT) in a patient with a dual-chamber implantable cardioverter-defibrillator and stable retrograde ventriculoatrial conduction time. The same rate-adaptive post-ventricular atrial refractory period (PVARP) duration had previously prevented PMT. Oversensing of atrial false signals from a defective lead shortened the PVARP with consequent sensing of retrograde conduction.