Microwave-assisted, hydrofluoric acid digestion is an increasingly common tool for the preparation of marine sediment samples for analysis by a variety of spectrometric techniques. Here we report that analysis of terrigenous-dominated sediment samples occasionally results in anomalously low values for several elements, including Al, Ba, Ca, Mg, and Sr. Measured concentrations of these elements increased with time between sample preparation and sample analysis, reaching stable values after 8-29 days. This lag is explained by the formation and subsequent dissolution of poorly soluble fluoride phases during digestion. Other elements, such as Fe, Mn, and Ti, showed little or no lag and were quickly measurable at a stable value. Full re-dissolution of the least soluble fluorides, which incorporate Al and Mg, requires up to four weeks at room temperature, and this duration can vary among sedimentary matrices. This waiting time can be reduced to 6 days (or shorter) if the samples are heated to ≈ 60°C for 24h.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22284480 DOI: 10.1016/j.talanta.2011.11.081