Verification of beam delivery using fibrosis after proton beam irradiation to the lung tumor.

Nobuyoshi Fukumitsu1, Yoshiko Oshiro, Takayuki Hashimoto, Toshiyuki Okumura, Masashi Mizumoto, Takashi Moritake, Koji Tsuboi, Takeji Sakae, Hideyuki Sakurai.   

Abstract

BACKGROUND AND
PURPOSE: To investigate the geometrical accuracy in proton beam therapy (PBT) to the lung by comparing the location of the fibrosis after PBT and the tumor using less invasive method. PATIENTS AND METHODS: We examined 50 lung tumors that had been treated by respiratory-gated PBT (33-74 Gray equivalents). Image registration and re-slicing of computed tomography (CT) before and after PBT were performed using an image analysis software system. We investigated whether the location of fibrosis was accurate with the tumor site and which factor among the respiratory rhythm, tumor volume, location, fixation, beam direction and clinical outcome was affected with accuracy.
RESULTS: The area of fibrosis was accurate with the tumor in 45 tumors (Group A) and not accurate in 5 tumors (Group B). All lesions with a small irregularity of respiratory rhythm showed fibrosis in an area accurate with the tumor. This percentage (100%) was higher than that of the lesions with a large irregularity of respiratory rhythm (67%, p = 0.037). Radiation pneumonitis was found for 1 lesion in group A and 2 lesions in Group B (p = 0.0009).
CONCLUSION: The geometrical accuracy of beam delivery to the lung was higher in the patients with a regular respiratory rhythm in PBT. The patients of Group B had greater risk of pneumonitis.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22281073     DOI: 10.1016/j.lungcan.2012.01.007

Source DB:  PubMed          Journal:  Lung Cancer        ISSN: 0169-5002            Impact factor:   5.705


  1 in total

1.  Reproducibility of image quality for moving objects using respiratory-gated computed tomography: a study using a phantom model.

Authors:  Nobuyoshi Fukumitsu; Masaya Ishida; Toshiyuki Terunuma; Masashi Mizumoto; Takayuki Hashimoto; Takashi Moritake; Toshiyuki Okumura; Takeji Sakae; Koji Tsuboi; Hideyuki Sakurai
Journal:  J Radiat Res       Date:  2012-09-10       Impact factor: 2.724

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.