Clinical experience with a new endobrochial blocker: the EZ-blocker.

Tamás Végh1, Marianna Juhász, Attila Enyedi, István Takács, József Kollár, Béla Fülesdi.   

Abstract

PURPOSE: In some thoracic surgical procedures, the insertion of a double-lumen tube (DLT) is not feasible, or the altered use of a DLT and a single-lumen tube (SLT) is justified during the surgery. In the present article we report our experience with a new bronchial blocker, the EZ-blocker, in clinical use.
METHODS: Data were obtained from ten patients undergoing thoracic surgery necessitating one-lung ventilation. For lung isolation, a combination of an SLT and an EZ-blocker was used. The time of insertion and positioning of the EZ-blocker, the lung deflation time with the EZ-blocker cuff inflated and deflated, and the cuff's minimal occlusion volume were recorded. Based on the CT scan, the diameter of the main bronchi and the angle of the bifurcation were measured offline.
RESULTS: The insertion duration of the EZ-blocker was 76 ± 15 s. Two malpositionings were caused by the too-deep positioning of the SLT used for introducing the EZ-blocker, which could be corrected within 65 ± 7 s. The use of the EZ-blocker allowed a short deflation time of the lung without (9.4 ± 0.7 s) and with (4.1 ± 0.7 s) administration of suctioning. The proper block was only dependent on the diameter of the main bronchi and was independent of the bifurcation angle.
CONCLUSIONS: Use of the EZ-blocker is easy and safe. The short insertion time and short lung deflation time through the lumen of the SLT allows its use in emergency situations or in cases of a difficult airway.

Entities:  

Mesh:

Year:  2012        PMID: 22278374     DOI: 10.1007/s00540-011-1315-0

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  10 in total

1.  Estimating the diameter of the left main bronchus.

Authors:  T Hampton; S Armstrong; W J Russell
Journal:  Anaesth Intensive Care       Date:  2000-10       Impact factor: 1.669

Review 2.  Indications and technique of Nuss procedure for pectus excavatum.

Authors:  Donald Nuss; Robert E Kelly
Journal:  Thorac Surg Clin       Date:  2010-11       Impact factor: 1.750

3.  CT assessment of tracheal carinal angle and its determinants.

Authors:  N Karabulut
Journal:  Br J Radiol       Date:  2005-09       Impact factor: 3.039

4.  Pressures exerted by endobronchial devices.

Authors:  Andrew Roscoe; George W Kanellakos; Karen McRae; Peter Slinger
Journal:  Anesth Analg       Date:  2007-03       Impact factor: 5.108

5.  Choosing a lung isolation device for thoracic surgery: a randomized trial of three bronchial blockers versus double-lumen tubes.

Authors:  Manu Narayanaswamy; Karen McRae; Peter Slinger; Geoffrey Dugas; George W Kanellakos; Andy Roscoe; Melanie Lacroix
Journal:  Anesth Analg       Date:  2009-04       Impact factor: 5.108

6.  Rapid pressure compensation by automated cuff pressure controllers worsens sealing in tracheal tubes.

Authors:  M Weiss; C Doell; N Koepfer; C Madjdpour; K Woitzek; V Bernet
Journal:  Br J Anaesth       Date:  2008-12-25       Impact factor: 9.166

7.  Lung isolation with a new Y-shaped endobronchial blocking device, the EZ-Blocker.

Authors:  H E Mungroop; P Tjong Yoe Wai; M N Morei; B G Loef; A H Epema
Journal:  Br J Anaesth       Date:  2010-01       Impact factor: 9.166

8.  Randomized clinical trial comparing double-lumen tube and EZ-Blocker for single-lung ventilation.

Authors:  K Ruetzler; G Grubhofer; W Schmid; D Papp; S Nabecker; D Hutschala; G Lang; H Hager
Journal:  Br J Anaesth       Date:  2011-04-14       Impact factor: 9.166

9.  Difficult tubes and difficult airways.

Authors:  J L Benumof
Journal:  J Cardiothorac Vasc Anesth       Date:  1998-04       Impact factor: 2.628

10.  The use of air in the inspired gas mixture during two-lung ventilation delays lung collapse during one-lung ventilation.

Authors:  Raynauld Ko; Karen McRae; Gail Darling; Thomas K Waddell; Desmond McGlade; Ken Cheung; Joel Katz; Peter Slinger
Journal:  Anesth Analg       Date:  2009-04       Impact factor: 5.108

  10 in total
  3 in total

1.  The EZ-blocker for one-lung ventilation in patients undergoing thoracic surgery: clinical applications and experience in 100 cases in a routine clinical setting.

Authors:  Andreas Moritz; Andrea Irouschek; Torsten Birkholz; Johannes Prottengeier; Horia Sirbu; Joachim Schmidt
Journal:  J Cardiothorac Surg       Date:  2018-06-25       Impact factor: 1.637

2.  Bronchial rupture following endobronchial blocker placement: a case report of a rare, unfortunate complication.

Authors:  Shuwen Oo; Rachel Hui Xuan Chia; Yue Li; Hari Kumar Sampath; Sophia Bee Leng Ang; Suresh Paranjothy; John Kit Chung Tam; Chang Chuan Melvin Lee
Journal:  BMC Anesthesiol       Date:  2021-08-30       Impact factor: 2.217

3.  Comparison of left double lumen tube and y-shaped and double-ended bronchial blocker for one lung ventilation in thoracic surgery-a randomised controlled clinical trial.

Authors:  Joachim Risse; Karsten Szeder; Ann-Kristin Schubert; Thomas Wiesmann; Hanns-Christian Dinges; Carsten Feldmann; Hinnerk Wulf; Karl Matteo Meggiolaro
Journal:  BMC Anesthesiol       Date:  2022-04-02       Impact factor: 2.217

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.