Literature DB >> 22276288

Microbiological contamination of spirometers - an exploratory study in general practice.

Kerry L Hancock1, Tjard R Schermer, Christine Holton, Alan J Crockett.   

Abstract

BACKGROUND: Spirometry is an important tool when diagnosing chronic respiratory conditions in general practice. However, the equipment may harbour pathogenic micro-organisms and cross-transmission of aerolised pathogens could occur if hygiene measures are insufficient.
METHODS: We assessed microbiological contamination in 16 spirometers from a convenience sample of South Australian general practices. Results We found potentially relevant microbiological contamination in three spirometers: two Pseudomonas spp.; one coagulase negative Staphylococcus sp. and one Alcaligenes sp. Although the three practices concerned all reported to have a written spirometer cleaning protocol in place, the frequency of spirometer disinfection did not match the manufacturers' DISCUSSION: Despite the small size of our study sample, we found potentially relevant microbiological contamination in 3 out of 16 spirometers from metropolitan general practices. The potential hazard of spirometers as reservoirs of microorganisms stresses the need for stricter attention to hygiene measures for spirometer maintenance in general practices.

Entities:  

Mesh:

Year:  2012        PMID: 22276288

Source DB:  PubMed          Journal:  Aust Fam Physician        ISSN: 0300-8495


  1 in total

1.  Increased constituent ratios of Klebsiella sp., Acinetobacter sp., and Streptococcus sp. and a decrease in microflora diversity may be indicators of ventilator-associated pneumonia: a prospective study in the respiratory tracts of neonates.

Authors:  Wei Lu; Jialin Yu; Qing Ai; Dong Liu; Chao Song; Luquan Li
Journal:  PLoS One       Date:  2014-02-20       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.