Literature DB >> 22269861

A comparative study on the reliability of an automated system for the evaluation of cell-based indirect immunofluorescence.

Alessandra Melegari1, Chiara Bonaguri, Annalisa Russo, Battistelli Luisita, Tommaso Trenti, Giuseppe Lippi.   

Abstract

BACKGROUND: Automated interpretations systems for anti-nuclear antibody (ANA), anti-double stranded DNA antibody (dsDNAab), and anti-neutrophil cytoplasmic antibody (ANCA) assessment by indirect immunofluorescence (IIF) have been recently introduced. The aim of this study was to compare the diagnostic performance of the automated IIF reading system AKLIDES with both traditional visual interpretation of IIF by laboratory experts and confirmatory tests.
METHODS: Visual and automated autoantibody interpretations of IIF findings using AKLIDES pattern recognition algorithms were performed for ANA on HEp-2 cells (n=182), dsDNAab on Crithidia luciliae (n=44) and ANCA on human neutrophils (n=46). All serum samples tested by IIF for ANCA and dsDNAab were also assessed with the corresponding enzyme-linked immunosorbent assays (ELISAs). Out of the 182 sera tested for ANA by IIF, 116 were also assessed for antibodies to extractable nuclear antigens (ENA) by ELISA and dot immunoassay (DIA).
RESULTS: ANA testing showed an excellent agreement between visual and AKLIDES reading (98.9%). The overall agreement of dsDNAab testing on C. luciliae substrate slides was 91.0%, whereas ANCA showed a concordance of 89.1%. There was a remarkable agreement of AKLIDES findings for dsDNAab with confirmatory tests.
CONCLUSION: Visual and automated interpretations of IIF findings for ANA, ANCA, and dsDNAab demonstrated a good agreement when assessing patients with suspected autoimmune diseases. Automated interpretation systems such AKLIDES may improve laboratory efficiency and support standardization of IIF in clinical laboratories.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22269861     DOI: 10.1016/j.autrev.2011.12.010

Source DB:  PubMed          Journal:  Autoimmun Rev        ISSN: 1568-9972            Impact factor:   9.754


  23 in total

1.  A fully automated IIF system for the detection of antinuclear antibodies and antineutrophil cytoplasmic antibodies.

Authors:  O Shovman; N Agmon-Levin; B Gilburd; T Martins; A Petzold; T Matthias; Y Shoenfeld
Journal:  Immunol Res       Date:  2015-02       Impact factor: 2.829

2.  Diagnostics and environmental factors.

Authors:  Elias Toubi
Journal:  Immunol Res       Date:  2015-02       Impact factor: 2.829

3.  Evaluation of an automated chemiluminescent immunoassay kit for antinuclear antibodies in autoimmune diseases.

Authors:  Yaron Zafrir; Boris Gilburd; Marina Garcia Carrasco; Shaye Kivity; María Sánchez-Castañón; Marcos López-Hoyos; Mathilda Mandel; Magdalena Szmyrka; Yehuda Shoenfeld; Nancy Agmon-Levin
Journal:  Immunol Res       Date:  2013-07       Impact factor: 2.829

4.  The burden of the variability introduced by the HEp-2 assay kit and the CAD system in ANA indirect immunofluorescence test.

Authors:  M Infantino; F Meacci; V Grossi; M Manfredi; M Benucci; M Merone; P Soda
Journal:  Immunol Res       Date:  2017-02       Impact factor: 2.829

Review 5.  Recent Approaches To Optimize Laboratory Assessment of Antinuclear Antibodies.

Authors:  Anne E Tebo
Journal:  Clin Vaccine Immunol       Date:  2017-12-05

Review 6.  Current and emerging techniques for ANCA detection in vasculitis.

Authors:  Elena Csernok; Frank Moosig
Journal:  Nat Rev Rheumatol       Date:  2014-06-03       Impact factor: 20.543

7.  Automatic reading of ANCA-slides: evaluation of the AKLIDES system.

Authors:  Jan Damoiseaux; Kathleen Mallet; Mia Vaessen; Jos Austen; Jan Willem Cohen Tervaert
Journal:  Clin Dev Immunol       Date:  2012-11-05

8.  Automated indirect immunofluorescence evaluation of antinuclear autoantibodies on HEp-2 cells.

Authors:  Jörn Voigt; Christopher Krause; Edda Rohwäder; Sandra Saschenbrecker; Melanie Hahn; Maick Danckwardt; Christian Feirer; Konstantin Ens; Kai Fechner; Erhardt Barth; Thomas Martinetz; Winfried Stöcker
Journal:  Clin Dev Immunol       Date:  2012-11-11

9.  Automated interpretation of ANCA patterns - a new approach in the serology of ANCA-associated vasculitis.

Authors:  Ilka Knütter; Rico Hiemann; Therese Brumma; Thomas Büttner; Kai Großmann; Marco Cusini; Francesca Pregnolato; Maria Orietta Borghi; Ursula Anderer; Karsten Conrad; Dirk Reinhold; Dirk Roggenbuck; Elena Csernok
Journal:  Arthritis Res Ther       Date:  2012-12-14       Impact factor: 5.156

Review 10.  New platform technology for comprehensive serological diagnostics of autoimmune diseases.

Authors:  Annika Willitzki; Rico Hiemann; Vanessa Peters; Ulrich Sack; Peter Schierack; Stefan Rödiger; Ursula Anderer; Karsten Conrad; Dimitrios P Bogdanos; Dirk Reinhold; Dirk Roggenbuck
Journal:  Clin Dev Immunol       Date:  2012-12-19
View more