SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging.

Tao Ai1, Abraham Padua, Frank Goerner, Mathias Nittka, Zbigniew Gugala, Siddharth Jadhav, Miguel Trelles, Raleigh F Johnson, Ronald W Lindsey, Xiaoming Li, Val M Runge.   

Abstract

OBJECTIVES: To evaluate the ability of four magnetic resonance imaging (MRI) techniques to correct for metallic artifacts. These techniques consisted of 3 2D techniques and one 3D technique. In 2D imaging the techniques View Angle Tilting (VAT), Slice Encoding for Metal Artifact Correction (SEMAC) and a technique that employed a combination of the first two (SEMAC-VAT) were evaluated. In 3D imaging the technique Multiple Slab acquisition with VAT based on a SPACE sequence was evaluated (MSVAT-SPACE).
MATERIALS AND METHODS: Agarose phantoms and tissue phantoms with two commonly used metal implants (stainless steel and titanium) as well as two volunteers with metal implants were imaged at 1.5T. All phantoms and volunteers were imaged using VAT, SEMAC, SEMAC-VAT and MSVAT-SPACE techniques, as well as 2D and 3D conventional imaging techniques. Each technique was optimized for different image contrast mechanisms. Artifact reduction was quantitatively assessed in the agarose phantoms by volumetric measurement. Image quality was qualitatively assessed by blinded reads employing two readers. Each reader independently viewed the tissue phantom images and in vivo human images. Statistical analysis was performed using a Friedman test, Wilcoxon test and weighted Cohen's kappa test.
RESULTS: T1-weighted, T2-weighted, PD-weighted and STIR image contrasts were successfully implemented with the evaluated artifact reduction sequences in both the phantom experiments and in vivo images. For all evaluated image contrasts and both metal implants, a reduction in the volume of metal artifacts was seen when compared with 2D conventional acquisitions. The 2D metal artifact volumes on average were reduced by 49% ± 16%, 56% ± 15% and 63% ± 15% for VAT, SEMAC and SEMAC-VAT acquisitions respectively. When Friedman and Wilcoxon tests were applied the difference in metal artifact volume was found to be statistically significant when VAT, SEMAC and SEMAC-VAT were compared with the 2D conventional techniques. In 3D imaging on average MSVAT-SPACE reduced metal artifact volume compared with the 3D conventional imaging technique by 72% ± 23% for all evaluated image contrasts and both metal implants. The metal artifact volume differences were statistically significant when MSVAT-SPACE was compared with the 3D conventional technique. The blinded reads demonstrated that SEMAC-VAT and MSVAT-SPACE had distinctly superior quality compared with conventional acquisitions. Quality was measured in terms of artifact size, distortions, image quality and visualization of bone marrow and soft tissues adjacent to metal implants. This was the case for both tissue phantom images and human images with good interobserver agreement.
CONCLUSIONS: SEMAC-VAT (2D) and MSVAT-SPACE (3D) demonstrated a consistent, marked reduction of metal artifacts for different metal implants and offered flexible image contrasts (T1, T2, PD and STIR) with high image quality. These techniques likely will improve the evaluation of postoperative patients with metal implants.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22266987     DOI: 10.1097/RLI.0b013e318240a919

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  20 in total

Review 1.  An illustrative review to understand and manage metal-induced artifacts in musculoskeletal MRI: a primer and updates.

Authors:  J P Dillenseger; S Molière; P Choquet; C Goetz; M Ehlinger; G Bierry
Journal:  Skeletal Radiol       Date:  2016-02-02       Impact factor: 2.199

2.  Usefulness of metal artifact reduction with WARP technique at 1.5 and 3T MRI in imaging metal-on-metal hip resurfacings.

Authors:  Andrea Lazik; Stefan Landgraeber; Patrick Schulte; Oliver Kraff; Thomas C Lauenstein; Jens M Theysohn
Journal:  Skeletal Radiol       Date:  2015-03-25       Impact factor: 2.199

3.  PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging.

Authors:  Tim Hilgenfeld; Marcel Prager; Alexander Heil; Franz Sebastian Schwindling; Mathias Nittka; David Grodzki; Peter Rammelsberg; Martin Bendszus; Sabine Heiland
Journal:  Eur Radiol       Date:  2017-07-11       Impact factor: 5.315

4.  Advanced metal artifact reduction MRI of metal-on-metal hip resurfacing arthroplasty implants: compressed sensing acceleration enables the time-neutral use of SEMAC.

Authors:  Jan Fritz; Benjamin Fritz; Gaurav K Thawait; Esther Raithel; Wesley D Gilson; Mathias Nittka; Michael A Mont
Journal:  Skeletal Radiol       Date:  2016-08-06       Impact factor: 2.199

5.  Unicompartmental knee arthroplasty MRI: impact of slice-encoding for metal artefact correction MRI on image quality, findings and therapy decision.

Authors:  Christoph A Agten; Filippo Del Grande; Sandro F Fucentese; Samuel Blatter; Christian W A Pfirrmann; Reto Sutter
Journal:  Eur Radiol       Date:  2015-01-28       Impact factor: 5.315

Review 6.  Recent advances in MRI of the head and neck, skull base and cranial nerves: new and evolving sequences, analyses and clinical applications.

Authors:  Philip Touska; Steve E J Connor
Journal:  Br J Radiol       Date:  2019-09-24       Impact factor: 3.039

7.  MR imaging near metallic implants using MAVRIC SL: initial clinical experience at 3T.

Authors:  Luis B Gutierrez; Bao H Do; Garry E Gold; Brian A Hargreaves; Kevin M Koch; Pauline W Worters; Kathryn J Stevens
Journal:  Acad Radiol       Date:  2014-11-27       Impact factor: 3.173

8.  Metal artifact reduction MRI of total ankle arthroplasty implants.

Authors:  Cesar de Cesar Netto; Lucas F Fonseca; Benjamin Fritz; Steven E Stern; Esther Raithel; Mathias Nittka; Lew C Schon; Jan Fritz
Journal:  Eur Radiol       Date:  2017-12-07       Impact factor: 5.315

9.  Hexagonal undersampling for faster MRI near metallic implants.

Authors:  Bragi Sveinsson; Pauline W Worters; Garry E Gold; Brian A Hargreaves
Journal:  Magn Reson Med       Date:  2014-02-18       Impact factor: 4.668

10.  In vitro assessment of knee MRI in the presence of metal implants comparing MAVRIC-SL and conventional fast spin echo sequences at 1.5 and 3 T field strength.

Authors:  Hans Liebl; Ursula Heilmeier; Sonia Lee; Lorenzo Nardo; Janina Patsch; Christopher Schuppert; Misung Han; Ina-Christine Rondak; Suchandrima Banerjee; Kevin Koch; Thomas M Link; Roland Krug
Journal:  J Magn Reson Imaging       Date:  2014-06-10       Impact factor: 4.813