Better stent expansion by two-time inflation of stent balloon and its responsible mechanism.

Yumiko Iwamoto1, Mitsunori Okamoto, Masaki Hashimoto, Yukihiro Fukuda, Akimichi Iwamoto, Toshitaka Iwasaki, Hiroki Kinoshita, Yasuki Kihara.   

Abstract

OBJECTIVES: We determined the effect of two-time inflation of the stent balloon on stent expansion and its responsible factor.
METHODS: Subjects included 61 patients with de novo coronary artery lesions, in whom 12 sirolimus-eluting, 27 paclitaxel-eluting, and 22 other stents were deployed twice at identical inflation pressures (11.3±2.3 atm) and inflation times (5, 10, 20, and 40 s). After the first and second deployments, minimum lumen diameter (MLD), minimum lumen area (MLA), and distensibility index (DI) were determined using intravascular ultrasound.
RESULTS: After the second inflation, MLA was significantly increased (5 s: 12.9%, 10 s: 14.5%, 20 s: 9.4%, 40 s: 9.5%). MLD and DI were also significantly increased. In the single and double inflation groups, DI in each group was significantly correlated with inflation time (single: r=0.409, double: r=0.351). DI was not significantly different between double 5-s and single 10-s inflations, between double 10-s and single 20-s inflations, or between double 20-s and single 40-s inflations. Additional stent balloon inflation by higher pressure in 30% and another balloon in 18% of the patients were required.
CONCLUSIONS: Two-time stent balloon inflation may allow better stent expansion regardless of inflation time and two-time inflation may be equivalent to longer inflation. Copyright Â
© 2012. Published by Elsevier Ltd.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22266460     DOI: 10.1016/j.jjcc.2011.12.003

Source DB:  PubMed          Journal:  J Cardiol        ISSN: 0914-5087            Impact factor:   3.159


  1 in total

1.  Impact of the Balloon Inflation Time and Pattern on the Coronary Stent Expansion.

Authors:  Jarosław Skowroński; Rafał Wolny; Jan Jastrzębski; Paweł Tyczyński; Karol Szlazak; Jerzy Pręgowski; Gary S Mintz; Karolina Liżewska; Wojciech Świeszkowski; Zbigniew Chmielak; Adam Witkowski
Journal:  J Interv Cardiol       Date:  2019-03-21       Impact factor: 2.279

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.