| Literature DB >> 22265562 |
Bang Lin Li1, Zhi Ling Wu, Chang Hong Xiong, Hong Qun Luo, Nian Bing Li.
Abstract
A carbon paste electrode modified with tin was used for the determination of trace cadmium by anodic stripping voltammetry. The electroanalytical performance for the determination of Cd(II) on the tin-coated carbon paste electrode(SnF-CPE) was better than that on the carbon paste electrode. The measuring conditions have been optimized. The measurement of trace cadmium on the SnF-CPE has the best response under the conditions of 0.10molL(-1) acetate buffer solution (pH 3.9), 3.5mgL(-1) Sn(II), deposition potential of -1.40V, and deposition time of 150s. The SnF-CPE revealed highly linear behavior in the concentration range of 2.0-90.0μgL(-1) with the detection limit of 1.13μgL(-1) for Cd(II). The developed sensor has been applied to the determination of Cd(II) in real water samples with satisfactory results.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22265562 DOI: 10.1016/j.talanta.2011.11.070
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057