Air-displacement plethysmography for the measurement of body composition in children aged 6-48 months.

Ryan P Rosendale1, Cynthia J Bartok.   

Abstract

INTRODUCTION: Air-displacement plethysmography (ADP) is an age-appropriate method for measuring relative fat mass (%FM) in children; however, the accuracy of this method has not been evaluated in children aged 5 y or younger.
RESULTS: Mean %FM values measured by ADP (17.9 ± 8.0%) and by total body water (TBW) (23.7 ± 6.3%) were significantly different (P < 0.001). Regression analysis of %FM by ADP vs. TBW provided a line of best fit with a slope of 0.089, r(2) = 0.013, and standard error of the estimate (SEE) = 6.3% FM (P = 0.40). DISCUSSION: The error was related to child weight and %FM, but not to behaviors (movement or vocalizations) occurring during the test sequence. A large portion of the error was attributable to imprecision in measuring small volumes. As currently designed, ADP is not an accurate method for measuring %FM in young children. Further investigation of the sources of variability will provide insight into ways of improving the accuracy of this technology for this population.
METHODS: This study examined the accuracy of an ADP system modified for young children (BOD POD; Life Measurement, Concord, CA) by comparing %FM results from ADP with those obtained from TBW by deuterium (D(2)O) dilution (reference method) in 72 children aged 6-48 mo.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22258086     DOI: 10.1038/pr.2011.42

Source DB:  PubMed          Journal:  Pediatr Res        ISSN: 0031-3998            Impact factor:   3.756


  3 in total

1.  Body composition of term healthy Indian newborns.

Authors:  V Jain; A V Kurpad; B Kumar; S Devi; V Sreenivas; V K Paul
Journal:  Eur J Clin Nutr       Date:  2015-09-16       Impact factor: 4.016

2.  Challenges in infant body composition.

Authors:  Ellen W Demerath; David A Fields
Journal:  Pediatr Res       Date:  2012-09       Impact factor: 3.756

Review 3.  Air displacement plethysmography (pea pod) in full-term and pre-term infants: a comprehensive review of accuracy, reproducibility, and practical challenges.

Authors:  Hajar Mazahery; Pamela R von Hurst; Christopher J D McKinlay; Barbara E Cormack; Cathryn A Conlon
Journal:  Matern Health Neonatol Perinatol       Date:  2018-06-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.