| Literature DB >> 22255003 |
Pamela T Bhatti1, Maxine A McClain.
Abstract
This paper describes a novel sensing approach for reducing power requirements of implantable vestibular prostheses. A passive, microfabricated polymeric inertial sensor for detecting angular head rotations based on the biomechanics of the human semicircular canal is described. Angular head motion is coded by deflection of a highly compliant capacitor plate placed in parallel with a rigid reference electrode. This capacitance change serves to detect instantaneous angular velocity along a given axis of rotation. Designed for integration with a microelectromechanical systems-based fully implantable vestibular prosthesis, this sensing method can provide substantial power savings when compared with contemporary gyroscopes.Entities:
Mesh:
Year: 2011 PMID: 22255003 DOI: 10.1109/IEMBS.2011.6090854
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X