Literature DB >> 22254798

Robust intraocular acquisition of the accommodation demand using eyeball movements.

Liane Rheinschmitt1, Fabian Ritter, Jörg A Nagel, Ulrich Gengenbach, Helmut Guth, Georg Bretthauer.   

Abstract

The Artificial Accommodation System is a mechatronic lens implant that will restore the ability of the human eye to accommodate. Therefore, the accommodation demand has to be acquired. One possibility is to measure the vergence angle of the eyeballs in reference to an external field. Using the earth magnetic field as reference the proof of this measuring principle was possible. Still there are drawbacks like high responsivity to interferences and limitations of the measuring range. The new approach is to use the gravity field as reference and thus reduce the responsivity to interferences. The measuring range can be expanded by combining both sensing principles.

Entities:  

Mesh:

Year:  2011        PMID: 22254798     DOI: 10.1109/IEMBS.2011.6090576

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Robust Design of an Optical Micromachine for an Ophthalmic Application.

Authors:  Ingo Sieber; Thomas Martin; Ulrich Gengenbach
Journal:  Micromachines (Basel)       Date:  2016-05-06       Impact factor: 2.891

  1 in total