| Literature DB >> 22254775 |
Hyejung Kim1, Chris Van Hoof, Refet Firat Yazicioglu.
Abstract
This paper describes a mixed-signal ECG processing platform with an 12-bit ADC architecture that can adapt its sampling rate according to the input signals rate of change. This enables the sampling of ECG signals with significantly reduced data rate without loss of information. The presented adaptive sampling scheme reduces the ADC power consumption, enables the processing of ECG signals with lower power consumption, and reduces the power consumption of the radio while streaming the ECG signals. The test results show that running a CWT-based R peak detection algorithm using the adaptively sampled ECG signals consumes only 45.6 μW and it leads to 36% less overall system power consumption.Mesh:
Year: 2011 PMID: 22254775 DOI: 10.1109/IEMBS.2011.6090414
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X