| Literature DB >> 22254628 |
Nicola J Cairns1, Clayton J Adam, Mark J Pearcy, James Smeathers.
Abstract
Non-invasive vibration analysis has been used extensively to monitor the progression of dental implant healing and stabilization. It is now being considered as a method to monitor femoral implants in transfemoral amputees. This paper evaluates two modal analysis excitation methods and investigates their capabilities in detecting changes at the interface between the implant and the bone that occur during osseointegration. Excitation of bone-implant physical models with the electromagnetic shaker provided higher coherence values and a greater number of modes over the same frequency range when compared to the impact hammer. Differences were detected in the natural frequencies and fundamental mode shape of the model when the fit of the implant was altered in the bone. The ability to detect changes in the model dynamic properties demonstrates the potential of modal analysis in this application and warrants further investigation.Mesh:
Year: 2011 PMID: 22254628 DOI: 10.1109/IEMBS.2011.6090463
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X