Literature DB >> 22254316

A new heating method with dielectric bolus using resonant cavity applicator for brain tumors.

Y Iseki1, K Kato, K Nakane, Y Shindo, K Tsuchiya, M Kubo, H Takahashi, T Uzuka, Y Fujii.   

Abstract

In this paper, we discuss a new method of controlling heating location in the proposed resonant cavity applicator. A dielectric bolus was used to non-invasively treat brain tumors. We have already confirmed that our heating system using resonant cavity is useful to non-invasively heat brain tumors. In order to heat tumors occurring at various locations, it is necessary to control the heating area with our heating system. First, we presented the proposed heating method and a phantom model to calculate temperature distributions. The results of temperature distributions were discussed. Second, a 3-D human head model constructed from 2-D MRI images was presented. The results of specific absorption rate distributions were discussed. From these results, it was found that the proposed heating method was useful to non-invasively treat brain tumors.

Entities:  

Mesh:

Year:  2011        PMID: 22254316     DOI: 10.1109/IEMBS.2011.6090086

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Deep heat therapy system with resonant cavity applicator for articular cartilage in knee osteoarthritis.

Authors:  Yasuhiro Shindo; Kenji Takahashi; Futoshi Ikuta; Yuya Iseki; Kazuhiro Harada; Kazuo Kato
Journal:  J Phys Ther Sci       Date:  2020-12-11