Evaluation of catheters and metallic catheter markers as calibration standard for measurement of coronary dimension.

W H Leung1, P A Demopulos, E L Alderman, W Sanders, M L Stadius.   

Abstract

Measurement of coronary dimension requires an accurate and reproducible dimensional reference. Angiographic catheters are frequently used for this purpose. We measured the angiographic diameters of a broad range of diagnostic and angioplasty guiding catheters by using two commonly used edge-detection algorithms. Angiographic diameters are significantly less than true catheter outer diameter. Therefore the use of contrast-filled catheters as a dimensional reference may lead to considerable error in vessel measurement with overestimation of absolute dimension. Tables of reference values for multiple catheter as calibration standard, tested under a variety of angiographic conditions. The metallic-tipped marker was found to have a better degree of reproducibility than catheters. These findings have implications for studies employing serial measurements of coronary artery dimension and for the clinical practice of estimating vessel diameter for choice of balloon size during angioplasty.

Mesh:

Year:  1990        PMID: 2225049     DOI: 10.1002/ccd.1810210305

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  4 in total

1.  The impact of vessel and catheter position on the measurement accuracy in catheter-based quantitative coronary angiography.

Authors:  W Wunderlich; B Roehrig; F Fischer; H R Arntz; R Agrawal; A Morguet; H P Schultheiss; D Horstkotte
Journal:  Int J Card Imaging       Date:  1998-08

2.  The QUinapril Ischemic Event Trial (QUIET) design and methods: evaluation of chronic ACE inhibitor therapy after coronary artery intervention.

Authors:  M Texter; R S Lees; B Pitt; R E Dinsmore; A C Uprichard
Journal:  Cardiovasc Drugs Ther       Date:  1993-04       Impact factor: 3.727

3.  Effect of calibration methods on the accuracy of angiographic measurements during transcatheter procedures in dogs.

Authors:  Lauren E Markovic; Brian A Scansen
Journal:  J Vet Intern Med       Date:  2018-03-12       Impact factor: 3.333

4.  Coronary artery size as a predictor of Y-graft patency following coronary artery bypass surgery.

Authors:  Danang Himawan Limanto; Hyoung Woo Chang; Dong Jung Kim; Jun Sung Kim; Kay-Hyun Park; Cheong Lim
Journal:  Medicine (Baltimore)       Date:  2021-01-15       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.