Literature DB >> 22246991

On the fluid-tissue contrast behavior of high-resolution steady-state sequences.

Oliver Bieri1, Carl Ganter, Klaus Scheffler.   

Abstract

In general, MR image contrast is expected to be resolution independent, but a pronounced loss of contrast is observed between fluids and tissues with contemporary musculoskeletal protocols (typical inplane resolution << 1 mm) using nonbalanced steady-state free precession, such as double echo steady state. For nonbalanced steady-state free precession, diffusion sensitivity increases with increasing spoiler moments which increase with decreasing voxel size, suggesting diffusion damping as the major cause for the observed contrast variation. This is confirmed by simulations and measurements indicating that for fluids, diffusion effects become apparent already for resolutions Δx < 1 mm, whereas tissues typically require Δx < 200 μm. Gradient spoiling, however, is generically not minimized but frequently applied along the readout direction. For anisotropic steady-state free precession scans, the loss of contrast between fluids and tissues from diffusion can thus be minimized by simply moving the spoiler gradients to the lowest resolution direction.
Copyright © 2012 Wiley Periodicals, Inc.

Mesh:

Year:  2012        PMID: 22246991     DOI: 10.1002/mrm.24154

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  3 in total

1.  Integrated SSFP for functional brain mapping at 7T with reduced susceptibility artifact.

Authors:  Kaibao Sun; Rong Xue; Peng Zhang; Zhentao Zuo; Zhongwei Chen; Bo Wang; Thomas Martin; Yi Wang; Lin Chen; Sheng He; Danny J J Wang
Journal:  J Magn Reson       Date:  2016-12-30       Impact factor: 2.229

2.  Utilizing shared information between gradient-spoiled and RF-spoiled steady-state MRI signals.

Authors:  Bragi Sveinsson; Garry E Gold; Brian A Hargreaves; Daehyun Yoon
Journal:  Phys Med Biol       Date:  2021-01-14       Impact factor: 3.609

3.  Phase-based fast 3D high-resolution quantitative T2 MRI in 7 T human brain imaging.

Authors:  Amir Seginer; Rita Schmidt
Journal:  Sci Rep       Date:  2022-08-18       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.