[Porous tantalum modular cups in revision hip arthroplasty].

Jacek Markuszewski1, Małgorzata Wierusz-Kozłowska, Waldemar Woźniak, Łukasz Łapaj, Paweł Kokoszka.   

Abstract

INTRODUCTION: Modem indications for use of cemented or uncemented implants in revision hip arthroplasty are comparable. The use of bone cement provides predictable short term results, but the lack of osteointegration deteriorates its long term durability. Uncemented implants undergo osteointegration, but reduced contact with host bone may cause early loosening. Modular porous tantalum implants are highly osteointegrative, and the availability of bone stock augments increases the versatility of this system.
MATERIAL AND METHODS: The study group consisted of 20 patients (21 hips) who underwent a revision hip arthroplasty using Trabecular Metal acetabular revision system. The follow-up period was between 7 and 31 months (average 20 months). All patients were evaluated with HHS and conventional radiography.
RESULTS: The mean HHS increased form40.45 (range 18.74-56.65) preoperatively to 78.8 (56.5-96) at the last follow-up. No mechanical failures, nor septic complications were noted.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22235641

Source DB:  PubMed          Journal:  Chir Narzadow Ruchu Ortop Pol


  2 in total

Review 1.  Systematic review on outcomes of acetabular revisions with highly-porous metals.

Authors:  Samik Banerjee; Kimona Issa; Bhaveen H Kapadia; Robert Pivec; Harpal S Khanuja; Michael A Mont
Journal:  Int Orthop       Date:  2013-11-01       Impact factor: 3.075

2.  Smoking may be a harbinger of early failure with ultraporous metal acetabular reconstruction.

Authors:  Adolph V Lombardi; Keith R Berend; Joanne B Adams; Ryan C Jefferson; Michael A Sneller
Journal:  Clin Orthop Relat Res       Date:  2013-02       Impact factor: 4.176

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.