Literature DB >> 22234022

Earthing defect: a cause for unstable hemodynamics.

R Barani Selvan1, Parnandi Bhaskar Rao, T R Ramachandran, David George Veliath.   

Abstract

Interference of monitored electrocardiogram (ECG) is a common event in intensive care units and operation theaters. Artifacts in the ECG tracing can resemble serious arrhythmia, thus leading to unnecessary usage of antiarrhythmics or electrical defibrillation. In addition, ECG artifacts may lead to serious hemodynamic consequences secondary to intra-aortic balloon pump (IABP) trigger failure. We report a rare event of IABP failure due to ECG artifact, which appeared on placement of the transthoracic echocardiography probe over the chest. Subsequent evaluation revealed potential current leakage from echocardiography machine secondary to earthing defect in the machine.

Entities:  

Mesh:

Year:  2012        PMID: 22234022     DOI: 10.4103/0971-9784.91482

Source DB:  PubMed          Journal:  Ann Card Anaesth        ISSN: 0971-9784


  3 in total

1.  Monitor screen display inversion: Potential for risk during anaesthesia.

Authors:  Pritee Bhirud; Sheetal Y Chiplonkar
Journal:  Indian J Anaesth       Date:  2014-05

2.  An unusual cause of electrocardiogram interference in the operation theater.

Authors:  Ankur Sharma; Renu Sinha
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2015 Jan-Mar

3.  Ultrasound-guided probe-generated artifacts stimulating ventricular tachycardia: A rare phenomenon.

Authors:  Rafat Shamim; Rudrashish Haldar; Ashutosh Kaushal
Journal:  Saudi J Anaesth       Date:  2017 Jan-Mar
  3 in total