Literature DB >> 22227082

Disagreement between ion selective electrode direct and indirect sodium measurements: estimation of the problem in a tertiary referral hospital.

Goce Dimeski1, Thomas J Morgan, Jeffrey J Presneill, Balasubramanian Venkatesh.   

Abstract

PURPOSE: We estimated the proportion of indirect ion selective electrode (ISE) plasma sodium analyses in intensive care unit (ICU) and hospital wide, exhibiting important disagreement with direct ISE results in relation to abnormal plasma protein concentrations.
MATERIALS AND METHODS: Direct and indirect ISE plasma sodium measurements were performed on 346 clinical specimens selected to reflect low, normal, or high total protein concentrations. Important intermethod disagreement was defined as |4| mmol/L or higher. Results were extrapolated to a 3-month laboratory series of 48,033 indirect ISE assays, including 2877 samples from intensive care.
RESULTS: Intermethod sodium disagreement at |4| mmol/L or higher was predicted for 25% of ICU samples. Almost all (97%) occurred in hypoproteinemic samples where indirect tended to exceed direct ISE estimates. Hospital wide, such disagreement was projected to occur in 8% of samples, of which the majority (70%) were also hypoproteinemic.
CONCLUSIONS: Important disagreement between indirect and direct ISE sodium measurements may exist in up to 1 in 4 ICU specimens and 1 in 12 hospital-wide samples. The main problem is indirect ISE overestimation associated with hypoproteinemia, potentially leading to misclassifications of pseudohypernatremia and pseudonormonatremia. We recommend that hospital laboratories consider standardization using direct ISE sodium measurement.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22227082     DOI: 10.1016/j.jcrc.2011.11.003

Source DB:  PubMed          Journal:  J Crit Care        ISSN: 0883-9441            Impact factor:   3.425


  19 in total

1.  Direct sodium measurement prevents underestimation of hyponatremia in critically ill patients.

Authors:  W van den Ancker; I A Haagen; P H J van der Voort
Journal:  Intensive Care Med       Date:  2015-02-12       Impact factor: 17.440

2.  Hyponatraemia in patients with multiple myeloma.

Authors:  Eitan Mirvis; Dunnya De-Silva; Atul Mehta
Journal:  BMJ Case Rep       Date:  2015-12-11

3.  Plasma sodium measurements by direct ion selective methods in laboratory and point of care may not be clinically interchangeable.

Authors:  Brigit A Weld; Thomas J Morgan; Jeffrey J Presneill; Steven Weier; David Cowley
Journal:  J Clin Monit Comput       Date:  2016-09-28       Impact factor: 2.502

Review 4.  Hyponatremia in patients with liver diseases: not just a cirrhosis-induced hemodynamic compromise.

Authors:  G Liamis; T D Filippatos; A Liontos; M S Elisaf
Journal:  Hepatol Int       Date:  2016-06-21       Impact factor: 6.047

5.  Prevalence of hyponatremia and association with mortality: results from NHANES.

Authors:  Sumit Mohan; Sue Gu; Amay Parikh; Jai Radhakrishnan
Journal:  Am J Med       Date:  2013-12       Impact factor: 4.965

6.  Association of unmeasured strong ions with outcome of hospitalized beef and dairy diarrheic calves.

Authors:  Diego E Gomez; Jeanne Lofstedt; Luis G Arroyo; Maureen Wichtel; Tammy Muirhead; Henri Stämpfli; J Trenton McClure
Journal:  Can Vet J       Date:  2017-10       Impact factor: 1.008

7.  Agreement of Potassium, Sodium, Glucose, and Hemoglobin Measured by Blood Gas Analyzer With Dry Chemistry Analyzer and Complete Blood Count Analyzer: A Two-Center Retrospective Analysis.

Authors:  Hongxiang Xie; Shiyu Lv; Sufeng Chen; Zhenzhen Pang; Deli Ye; Jianzhuang Guo; Wanju Xu; Weidong Jin
Journal:  Front Med (Lausanne)       Date:  2022-04-01

8.  Comparison of different methods for measurement of electrolytes in patients admitted to the intensive care unit.

Authors:  Sinan Yilmaz; Hilal B Uysal; Mücahit Avcil; Mustafa Yilmaz; Bekir Dağlı; Murat Bakış; Imran K Ömürlü
Journal:  Saudi Med J       Date:  2016-03       Impact factor: 1.484

9.  Are sodium and potassium results on arterial blood gas analyzer equivalent to those on electrolyte analyzer?

Authors:  Shalini Gupta; Ashwani K Gupta; Kamaljit Singh; Minni Verma
Journal:  Indian J Crit Care Med       Date:  2016-04

10.  Diagnostic accuracy of calculated serum osmolarity to predict dehydration in older people: adding value to pathology laboratory reports.

Authors:  Lee Hooper; Asmaa Abdelhamid; Adam Ali; Diane K Bunn; Amy Jennings; W Garry John; Susan Kerry; Gregor Lindner; Carmen A Pfortmueller; Fredrik Sjöstrand; Neil P Walsh; Susan J Fairweather-Tait; John F Potter; Paul R Hunter; Lee Shepstone
Journal:  BMJ Open       Date:  2015-10-21       Impact factor: 2.692

View more