| Literature DB >> 22226410 |
Panpan Wang1, Lei Ge, Mei Yan, Xianrang Song, Shenguang Ge, Jinghua Yu.
Abstract
In this study, electrochemical immunoassay was introduced into the recently proposed microfluidic paper-based analytical device (μPADs). To improve the performance of electrochemical immunoassay on μPAD for point-of-care testing (POCT), a novel wax-patterned microfluidic paper-based three-dimensional electrochemical device (3D-μPED) was demonstrated based on the multi-walled carbon nanotubes (MWCNTs) modified μPAD. Using typical HRP-O-Phenylenediamine-H(2)O(2) electrochemical system, a sandwich immunoassay on this 3D-μPED for sensitive diagnosis of two tumor markers simultaneously in real clinical serum samples was developed with a linear range of 0.001-75.0 UmL(-1) for cancer antigen 125 and 0.05-50.0 ngmL(-1) for carcinoembryonic antigen. In addition, this 3D-μPED can be easily integrated and combined with the recently emerging paper electronics to further develop simple, sensitive, low-cost, disposable and portable μPAD for POCT, public health and environmental monitoring in remote regions, developing or developed countries.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22226410 DOI: 10.1016/j.bios.2011.12.021
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618