Literature DB >> 22226258

Generic safety requirements for developing safe insulin pump software.

Yi Zhang1, Raoul Jetley, Paul L Jones, Arnab Ray.   

Abstract

BACKGROUND: The authors previously introduced a highly abstract generic insulin infusion pump (GIIP) model that identified common features and hazards shared by most insulin pumps on the market. The aim of this article is to extend our previous work on the GIIP model by articulating safety requirements that address the identified GIIP hazards. These safety requirements can be validated by manufacturers, and may ultimately serve as a safety reference for insulin pump software. Together, these two publications can serve as a basis for discussing insulin pump safety in the diabetes community.
METHODS: In our previous work, we established a generic insulin pump architecture that abstracts functions common to many insulin pumps currently on the market and near-future pump designs. We then carried out a preliminary hazard analysis based on this architecture that included consultations with many domain experts. Further consultation with domain experts resulted in the safety requirements used in the modeling work presented in this article.
RESULTS: Generic safety requirements for the GIIP model are presented, as appropriate, in parameterized format to accommodate clinical practices or specific insulin pump criteria important to safe device performance.
CONCLUSIONS: We believe that there is considerable value in having the diabetes, academic, and manufacturing communities consider and discuss these generic safety requirements. We hope that the communities will extend and revise them, make them more representative and comprehensive, experiment with them, and use them as a means for assessing the safety of insulin pump software designs. One potential use of these requirements is to integrate them into model-based engineering (MBE) software development methods. We believe, based on our experiences, that implementing safety requirements using MBE methods holds promise in reducing design/implementation flaws in insulin pump development and evolutionary processes, therefore improving overall safety of insulin pump software.
© 2011 Diabetes Technology Society.

Entities:  

Mesh:

Year:  2011        PMID: 22226258      PMCID: PMC3262707          DOI: 10.1177/193229681100500612

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  2 in total

1.  A hazard analysis for a generic insulin infusion pump.

Authors:  Yi Zhang; Paul L Jones; Raoul Jetley
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

2.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

  2 in total
  3 in total

1.  Evaluation of Intradermal and Subcutaneous Infusion Set Performance Under 24-Hour Basal and Bolus Conditions.

Authors:  Elaine McVey; Steven Keith; Joshua K Herr; Diane Sutter; Ronald J Pettis
Journal:  J Diabetes Sci Technol       Date:  2015-08-27

Review 2.  Insulin Pump Therapy for Patients With Type 2 Diabetes Mellitus: Evidence, Current Barriers, and New Technologies.

Authors:  Guido Freckmann; Sina Buck; Delia Waldenmaier; Bernhard Kulzer; Oliver Schnell; Ulrich Gelchsheimer; Ralph Ziegler; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2020-06-01

3.  DUALarm: An open-source and 3D-printable device for upper limb neurorehabilitation.

Authors:  Tito Dinon; Marco Caimmi; Andrea Chiavenna; Matteo Malosio; Alessio Prini; Alessandro Scano; Lorenzo Molinari Tosatti; Cristian Currò; Bruno Lenzi; Valentino Megale
Journal:  J Rehabil Assist Technol Eng       Date:  2018-01-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.