A theoretical comparison of x-ray angiographic image quality using energy-dependent and conventional subtraction methods.

Jesse Tanguay1, Ho Kyung Kim, Ian A Cunningham.   

Abstract

PURPOSE: X-ray digital subtraction angiography (DSA) is widely used for vascular imaging. However, the need to subtract a mask image can result in motion artifacts and compromised image quality. The current interest in energy-resolving photon-counting (EPC) detectors offers the promise of eliminating motion artifacts and other advanced applications using a single exposure. The authors describe a method of assessing the iodine signal-to-noise ratio (SNR) that may be achieved with energy-resolved angiography (ERA) to enable a direct comparison with other approaches including DSA and dual-energy angiography for the same patient exposure.
METHODS: A linearized noise-propagation approach, combined with linear expressions of dual-energy and energy-resolved imaging, is used to describe the iodine SNR. The results were validated by a Monte Carlo calculation for all three approaches and compared visually for dual-energy and DSA imaging using a simple angiographic phantom with a CsI-based flat-panel detector.
RESULTS: The linearized SNR calculations show excellent agreement with Monte Carlo results. While dual-energy methods require an increased tube heat load of 2× to 4× compared to DSA, and photon-counting detectors are not yet ready for angiographic imaging, the available iodine SNR for both methods as tested is within 10% of that of conventional DSA for the same patient exposure over a wide range of patient thicknesses and iodine concentrations.
CONCLUSIONS: While the energy-based methods are not necessarily optimized and further improvements are likely, the linearized noise-propagation analysis provides the theoretical framework of a level playing field for optimization studies and comparison with conventional DSA. It is concluded that both dual-energy and photon-counting approaches have the potential to provide similar angiographic image quality to DSA.

Entities:  

Mesh:

Year:  2012        PMID: 22225282     DOI: 10.1118/1.3658728

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Theoretical and Monte Carlo optimization of a stacked three-layer flat-panel x-ray imager for applications in multi-spectral medical imaging.

Authors:  Sebastian Lopez Maurino; Aldo Badano; Ian A Cunningham; Karim S Karim
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

2.  New infusion device for use in acquisition of images during endovascular procedures: an experimental model.

Authors:  Bruno Pagnin Schmid; Isadora Cardoso de Alencar; Carolina Masson; Giovani José Dal Poggetto Molinari; Fábio Hüsemann Menezes
Journal:  J Vasc Bras       Date:  2021-06-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.