Literature DB >> 22225271

A forward bias method for lag correction of an a-Si flat panel detector.

Jared Starman1, Carlo Tognina, Larry Partain, Rebecca Fahrig.   

Abstract

PURPOSE: Digital a-Si flat panel (FP) x-ray detectors can exhibit detector lag, or residual signal, of several percent that can cause ghosting in projection images or severe shading artifacts, known as the radar artifact, in cone-beam computed tomography (CBCT) reconstructions. A major contributor to detector lag is believed to be defect states, or traps, in the a-Si layer of the FP. Software methods to characterize and correct for the detector lag exist, but they may make assumptions such as system linearity and time invariance, which may not be true. The purpose of this work is to investigate a new hardware based method to reduce lag in an a-Si FP and to evaluate its effectiveness at removing shading artifacts in CBCT reconstructions. The feasibility of a novel, partially hardware based solution is also examined.
METHODS: The proposed hardware solution for lag reduction requires only a minor change to the FP. For pulsed irradiation, the proposed method inserts a new operation step between the readout and data collection stages. During this new stage the photodiode is operated in a forward bias mode, which fills the defect states with charge. A Varian 4030CB panel was modified to allow for operation in the forward bias mode. The contrast of residual lag ghosts was measured for lag frames 2 and 100 after irradiation ceased for standard and forward bias modes. Detector step response, lag, SNR, modulation transfer function (MTF), and detective quantum efficiency (DQE) measurements were made with standard and forward bias firmware. CBCT data of pelvic and head phantoms were also collected.
RESULTS: Overall, the 2nd and 100th detector lag frame residual signals were reduced 70%-88% using the new method. SNR, MTF, and DQE measurements show a small decrease in collected signal and a small increase in noise. The forward bias hardware successfully reduced the radar artifact in the CBCT reconstruction of the pelvic and head phantoms by 48%-81%.
CONCLUSIONS: Overall, the forward bias method has been found to greatly reduce detector lag ghosts in projection data and the radar artifact in CBCT reconstructions. The method is limited to improvements of the a-Si photodiode response only. A future hybrid mode may overcome any limitations of this method.

Entities:  

Mesh:

Year:  2012        PMID: 22225271      PMCID: PMC3257750          DOI: 10.1118/1.3664004

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-08       Impact factor: 4.071

2.  Cone-beam computed tomography with a flat-panel imager: effects of image lag.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

3.  An empirical method for lag correction in cone-beam CT.

Authors:  N Mail; D J Moseley; J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

4.  Investigation into the optimal linear time-invariant lag correction for radar artifact removal.

Authors:  Jared Starman; Josh Star-Lack; Gary Virshup; Edward Shapiro; Rebecca Fahrig
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

  4 in total
  1 in total

1.  A nonlinear lag correction algorithm for a-Si flat-panel x-ray detectors.

Authors:  Jared Starman; Josh Star-Lack; Gary Virshup; Edward Shapiro; Rebecca Fahrig
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.