Literature DB >> 22225193

Junction temperature measurement of light emitting diode by electroluminescence.

S M He1, X D Luo, B Zhang, L Fu, L W Cheng, J B Wang, W Lu.   

Abstract

Junction temperature (JT) is a key parameter of the performance and lifetime of light emitting diodes (LEDs). In this paper, a mobile instrument system has been developed for the non-contact measurement of JTs of LED under LabVIEW control. The electroluminescence (EL) peak shift of the LED is explored to measure the JT. Commercially available high power blue LEDs are measured. A linear relation between emission peak shift and JT is found. The accuracy of the JT is about 1 °C determined by the precision of the emission peak shift, ±0.03 nm, at 3σ standard deviation for blue LED. Using this system, on-line temperature rise curves of LED lamps are determined.

Year:  2011        PMID: 22225193     DOI: 10.1063/1.3664619

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Noncontact and instant detection of phosphor temperature in phosphor-converted white LEDs.

Authors:  Tsung-Hsun Yang; Hsu-Yi Huang; Ching-Cherng Sun; Benoît Glorieux; Xuan-Hao Lee; Yeh-Wei Yu; Te-Yuan Chung
Journal:  Sci Rep       Date:  2018-01-10       Impact factor: 4.379