| Literature DB >> 22223467 |
Jeannine M Conway1, Angela K Birnbaum, Susan E Marino, James C Cloyd, Rory P Remmel.
Abstract
Topiramate (Topamax®) is an antiepileptic medication used as adjunctive and monotherapy in patients with epilepsy and for migraine prophylaxis. A GC-MS assay was developed that was capable of detecting topiramate plasma concentrations following a single rectal or oral dose administration. Topiramate plasma samples were prepared by solid-phase extraction and were quantified by GC-MS analysis. The topiramate standard curves were split from 0.1 to 4 µg/mL and from 4 to 40 µg/mL in order to give a more accurate determination of the topiramate concentration. The accuracy of the standards ranged from 94.6 to 107.3% and the precision (%CV) ranged from 1.0 to 5.3% for both curves at all concentrations. The %CV for quality controls was <7.6%. The assay is both accurate and precise and will be used to complete future pharmacokinetic studies.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22223467 PMCID: PMC3776594 DOI: 10.1002/bmc.1749