Literature DB >> 22220569

Precise pH measuring of platelet concentrates containing additive solution--the impact of the temperature.

J Ringwald1, R Luther, R Zimmermann, J Strobel, D Weiss, R Eckstein.   

Abstract

BACKGROUND AND OBJECTIVES: Blood gas analysers measuring pH at 37°C (pH37) are widely used for pH determination of platelet (PLT) concentrates (PCs). For reporting pH at 22°C (pH22), converting of pH37 using the correct conversion factor is mandatory. For PCs stored in PLT additive solution (PAS), such conversion factors are not yet widely available. We studied pH in samples of PCs with different PAS/plasma ratios during warming from 22 to 37°C.
MATERIALS AND METHODS: We measured pH in 39 samples containing modified PAS-III (PAS-IIIM) with a plasma carryover of 20%, 30% or 40% or no PAS-IIIM. Differences between pH22 and pH37 (dpH) were compared within and between study groups. Correlation between pH22 and dpH was tested. Additional measurements in 33 samples with three different PLT counts were performed to study the influence of PLT count on dpH.
RESULTS: pH22 and pH37 within each group and dpH or dpH/dT between study groups differed significantly. The dpH was 0·135 ± 0·040, 0·021 ± 0·009, 0·033 ± 0·011 and 0·048 ± 0·017 for samples containing 100%, 20%, 30% or 40% plasma, respectively. Correlation between dpH and pH22 was strong in 100% (r = 0·696, P < 0·001), weaker in 30% and 40% (r = 0·367, P = 0·022 and r = 0·345, P = 0·032, respectively) and not existing in 20% plasma (r = 0·153, P = 0·354). PLT count did not influence the dpH significantly.
CONCLUSION: The dpH is dependent on different PAS-IIIM/plasma ratios and pH range. For precise reporting of pH22, the respective dpH must be used if converting is necessary. Preferably, the pH should be reported at 37°C or measured directly at 22°C.
© 2012 The Author(s). Vox Sanguinis © 2012 International Society of Blood Transfusion.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22220569     DOI: 10.1111/j.1423-0410.2011.01578.x

Source DB:  PubMed          Journal:  Vox Sang        ISSN: 0042-9007            Impact factor:   2.144


  2 in total

1.  A Comparative Assessment of Quality of Platelet Concentrates Prepared by Buffy Coat Poor Platelet Concentrate Method and Apheresis Derived Platelet Concentrate Method.

Authors:  R S Mallhi; Sudeep Kumar; Joseph Philip
Journal:  Indian J Hematol Blood Transfus       Date:  2014-12-03       Impact factor: 0.900

2.  Haemostatic profile of riboflavin-treated apheresis platelet concentrates.

Authors:  Eleni Petrou; Georgios K Nikolopoulos; Anastasios G Kriebardis; Katerina Pantavou; Electra Loukopoulou; Andreas G Tsantes; Hara T Georgatzakou; Eirini Maratou; Evdoxia Rapti; Sofia Mellou; Styliani Kokoris; Argyri Gialeraki; Argirios E Tsantes
Journal:  Blood Transfus       Date:  2021-05-21       Impact factor: 5.752

  2 in total