Automated office versus home measurement of blood pressure in the assessment of morning hypertension.

Emmanuel A Andreadis1, Epameinondas T Angelopoulos, Athanasios P Tsakanikas, Gerasimos D Agaliotis, Sofia D Kravvariti, George P Mousoulis.   

Abstract

OBJECTIVE: To compare the quality and accuracy of morning blood pressure (BP) readings as taken by automated office BP (AOBP) and morning home BP (mHBP) techniques using morning ambulatory BP (mABP) measurements as the gold standard.
METHODS: A total of 139 individuals were included, 70 men and 69 women, mean age 53±13 years. The average AOBP readings as measured using a Microlife Watch BP office device taking triplicate automated simultaneous readings of both arms were compared with mHBP monitored on 6 routine days, using a validated automated electronic device. Both modalities were also compared with the ambulatory readings of the 3 h of waking (mABP3h).
RESULTS: The AOBP values were slightly higher than the mABP3h (mean difference 8.2 mmHg, 95% limits of agreement, -18.8 to 35.2 mmHg for the systolic BP and mean difference 4.3 mmHg, 95% limits of agreement, -15.3 to 23.9 mmHg for the diastolic BP). Systolic and diastolic AOBP readings correlated with mABP3h (r=0.66, P=0.001 and r=0.64, P=0.001, respectively). Agreement was fair between AOBP and mHBP in the detection of morning hypertensive patients (agreement 70%, κ=0.32) as compared with AOBP and mABP3h (agreement 67%, κ=0.32) and mHBP and mABP3h (agreement 65%, κ=0.31).
CONCLUSION: The AOBP technique could replace mHBP monitoring in the assessment of morning BP, as it provides comparable data in relation to the awake ambulatory BP. Given the simplicity of this method, it could be more readily applied in a larger number of individuals.

Entities:  

Mesh:

Year:  2012        PMID: 22218221     DOI: 10.1097/MBP.0b013e3283503760

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  4 in total

1.  Comparing Automated Office Blood Pressure Readings With Other Methods of Blood Pressure Measurement for Identifying Patients With Possible Hypertension: A Systematic Review and Meta-analysis.

Authors:  Michael Roerecke; Janusz Kaczorowski; Martin G Myers
Journal:  JAMA Intern Med       Date:  2019-03-01       Impact factor: 44.409

2.  Prospective blood pressure measurement in renal transplant recipients.

Authors:  V G David; B Yadav; L Jeyaseelan; M N Deborah; S Jacob; S Alexander; S Varughese; G T John
Journal:  Indian J Nephrol       Date:  2014-05

3.  Attended and Unattended Automated Office Blood Pressure Measurements Have Better Agreement With Ambulatory Monitoring Than Conventional Office Readings.

Authors:  Emmanuel A Andreadis; Charalampia V Geladari; Epameinondas T Angelopoulos; Florentia S Savva; Anna I Georgantoni; Vasilios Papademetriou
Journal:  J Am Heart Assoc       Date:  2018-04-07       Impact factor: 5.501

4.  Blood pressure measurement: Should technique define targets?

Authors:  Swapnil Hiremath; Tim Ramsay; Marcel Ruzicka
Journal:  J Clin Hypertens (Greenwich)       Date:  2021-07-16       Impact factor: 3.738

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.