Abstract

BACKGROUND: Transarticular screw fixation is seen as the "gold standard" in instrumented fusion of C1 and C2. However, drawbacks are the necessity of a reduction before instrumentation and a risk of vertebral artery injury. Therefore, C1 lateral mass and C2 isthmic screws are an alternative. The present study assessed the feasibility of C1-2 stabilization with C1 lateral mass and C2 isthmic screws and evaluated quality of life.
METHOD: All data of 35 consecutive patients treated from May 2006 to September 2009 were collected. Patients had C1 lateral mass and C2 isthmic screws.
RESULTS: Twenty patients were operated on for traumatic instabilities, six for neoplastic instabilities, five for infectious instabilities and two each for degenerative and congenital instabilities. Sixty-six of 70 C1 screws had an ideal position, while four were placed suboptimal without the need for revision. Twelve of 68 C2 screws were not ideal but acceptable; one screw needed a surgical revision. There was one non-surgery related case of neurological deterioration after multilevel instrumentation. No vascular injuries occurred. Realignment was correct in all patients. After a median follow-up of 12 months, patients showed a reduction of pain, disability and improvements in EQ-5D items. SF36 data compared with a normative population and a historical cohort showed lower levels of function in all domains.
CONCLUSION: C1-C2 instrumented fusion with lateral mass and isthmic screws is a safe procedure. Sufficient screw position and alignment was possible in all cases. Therefore, at our institution transarticular screws were abandoned in favor of C1 lateral mass and C2 isthmic screws.

Entities:  

Mesh:

Year:  2011        PMID: 22200812     DOI: 10.1007/s00701-011-1224-x

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  5 in total

1.  Biomechanical comparison of a novel C1 posterior U-construct with four other techniques in a C1-C2 fixation model.

Authors:  Joshua P Herzog; Nicholas J Zarkadis; Gautham Prabhakar; Nicholas A Kusnezov
Journal:  J Orthop       Date:  2018-05-18

2.  Individualized 3D printed navigation template-assisted atlantoaxial pedicle screws vs. free-hand screws for the treatment of upper cervical fractures.

Authors:  Guoqi Niu; Jiawei Cheng; Lutan Liu; Chao Li; Gong Zhou; Hui Chen; Tao Liu; Hu Nie; Zheng Sun; Weili Jiang; Qiankun Zhou; Baoyin Zhao; Jun Zhu; Ruochen Yu; Yalong Guo; Yi Yang; Jianzhong Bai
Journal:  Front Surg       Date:  2022-09-26

3.  An anatomic consideration of C2 vertebrae artery groove variation for individual screw implantation in axis.

Authors:  Janhua Wang; Hong Xia; Qingshui Ying; Yang Lu; Zenghui Wu; Fzhi Ai; Xiangyang Ma
Journal:  Eur Spine J       Date:  2013-05-10       Impact factor: 3.134

4.  Repeated complication following atlantoaxial fusion: a case report.

Authors:  Chang Hyun Oh; Gyu Yeul Ji; Hyun Sung Seo; Seung Hwan Yoon; Dongkeun Hyun; Hyeong-Chun Park
Journal:  Korean J Spine       Date:  2014-03-31

5.  Accuracy and safety of C2 pedicle or pars screw placement: a systematic review and meta-analysis.

Authors:  Parisa Azimi; Taravat Yazdanian; Edward C Benzel; Hossein Nayeb Aghaei; Shirzad Azhari; Sohrab Sadeghi; Ali Montazeri
Journal:  J Orthop Surg Res       Date:  2020-07-20       Impact factor: 2.359

  5 in total