Literature DB >> 22198353

Effect of graded facetectomy on biomechanics of Dynesys dynamic stabilization system.

Ali Kiapour1, Divya Ambati, Robert W Hoy, Vijay K Goel.   

Abstract

STUDY
DESIGN: Finite element (FE) method was used to compare the biomechanics of L3-S1 lumbar spine with graded facetectomy before and after placement of Dynesys.
OBJECTIVE: To evaluate the biomechanics of Dynesys as a function of graded bilateral facetectomies. SUMMARY OF BACKGROUND DATA: Spinal fusion or posterior dynamic stabilization systems are used to restore stability after facetectomies.
METHODS: The intact FE spine was modified to simulate decompression at L4-L5 with 50% and 75% and total facetectomy with/without dynamic stabilization with Dynesys. Biomechanics of the implanted level was investigated under different physiological loadings.
RESULTS: Total facetectomy increased the motion in extension (8.7° vs. 2.7° for intact) and axial rotation (8.4° vs. 2.4° for intact). However the decrease in motion in the Dynesys model ranged from 65% in axial rotation to 80% in flexion for all facetectomies, except in the total facetectomy axial rotation case (motion higher than intact). The center of rotation of dynamic stabilized segment moved inferior/posterior in partial facetectomy and superior/posterior in total facetectomy with respect to the intact and destabilized cases. The Dynesys screws observed peak stresses up to 28% higher than those of a rigid fixation system in certain loadings, such as lateral bending and extension. The critical loosening torque applied to the screws in total facetectomy case was 6 times the partial facetectomy case in axial rotation.
CONCLUSION: Partial facetectomy had a minimal effect on range of motion on the Dynesys-implanted segment. However, in the case of total facetectomy the motion increased by almost 40% in flexion and by 200% in axial rotation. The higher stresses applied to the screws in Dynesys in specific loadings may lead to higher risk of screw failure in Dynesys than in a generic rigid fixation construct.

Entities:  

Mesh:

Year:  2012        PMID: 22198353     DOI: 10.1097/BRS.0b013e3182463775

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  13 in total

1.  Elastic resistance of the spine: Why does motion preservation surgery almost fail?

Authors:  Alessandro Landi
Journal:  World J Clin Cases       Date:  2013-07-16       Impact factor: 1.337

2.  Biomechanical analysis and design of a dynamic spinal fixator using topology optimization: a finite element analysis.

Authors:  Hung-Ming Lin; Chien-Lin Liu; Yung-Ning Pan; Chang-Hung Huang; Shih-Liang Shih; Shun-Hwa Wei; Chen-Sheng Chen
Journal:  Med Biol Eng Comput       Date:  2014-04-16       Impact factor: 2.602

3.  Simulation of L-4 lumbar spine model of motorist exposed to vibration from speed hump.

Authors:  B Jain A R Tony; M S Alphin; G Sri Krishnan
Journal:  J Orthop       Date:  2020-08-27

4.  Biomechanical Comparison between Isobar and Dynamic-Transitional Optima (DTO) Hybrid Lumbar Fixators: A Lumbosacral Finite Element and Intersegmental Motion Analysis.

Authors:  Shih-Hao Chen; Chih-Kun Hsiao; Chih-Wei Wang; Hsiang-Ho Chen; Zheng-Cheng Zhong
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

5.  Preoperative facet joint arthropathy does not impact long-term clinical outcomes after lumbar-stability-preserving decompression and dynesys fixation.

Authors:  Po-Hsin Chou; Hsi-Hsien Lin; Yu-Cheng Yao; Shih-Tien Wang; Ming-Chau Chang; Chien-Lin Liu
Journal:  Sci Rep       Date:  2021-05-28       Impact factor: 4.379

6.  Biomechanics of disc degeneration.

Authors:  V Palepu; M Kodigudla; V K Goel
Journal:  Adv Orthop       Date:  2012-06-17

7.  On the Use of Biaxial Properties in Modeling Annulus as a Holzapfel-Gasser-Ogden Material.

Authors:  Narjes Momeni Shahraki; Ali Fatemi; Vijay K Goel; Anand Agarwal
Journal:  Front Bioeng Biotechnol       Date:  2015-06-03

8.  Effect of Graded Facetectomy on Lumbar Biomechanics.

Authors:  Zhi-Li Zeng; Rui Zhu; Yang-Chun Wu; Wei Zuo; Yan Yu; Jian-Jie Wang; Li-Ming Cheng
Journal:  J Healthc Eng       Date:  2017-02-19       Impact factor: 2.682

9.  Effects of cord pretension and stiffness of the Dynesys system spacer on the biomechanics of spinal decompression- a finite element study.

Authors:  Shih-Liang Shih; Chien-Lin Liu; Li-Ying Huang; Chang-Hung Huang; Chen-Sheng Chen
Journal:  BMC Musculoskelet Disord       Date:  2013-06-19       Impact factor: 2.362

10.  Clinical and Radiological Outcomes of Foraminal Decompression Using Unilateral Biportal Endoscopic Spine Surgery for Lumbar Foraminal Stenosis.

Authors:  Ju-Eun Kim; Dae-Jung Choi; Eugene J Park
Journal:  Clin Orthop Surg       Date:  2018-11-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.