Contact pressure on ACL hamstring grafts in the bone tunnel with interference screw fixation--dynamic adaptation under load.

Dominik C Meyer1, Michael Stalder, Peter P Koch, Jess G Snedeker, Mazda Farshad.   

Abstract

INTRODUCTION: Interference screws used in fixation of anterior cruciate ligament (ACL) hamstring grafts create mechanical hold by forcing the graft into frictional contact with the bone tunnel. We analyzed the resultant graft-tunnel contact pressure using an in vitro model of human cadaver 8mm hamstring grafts.
METHODS: Contact characteristics were assessed using both pressure sensitive films and a force sensor. Two screw sizes were investigated (8 and 9 mm in an 8mm Sawbone tunnel), both with and without a bone wedge between graft and screw. Separately, time dependent relaxation of contact force was recorded over a one hour epoch and associated tendon water loss was measured. Pullout testing of 8mm tendon grafts from 8mm holes in Sawbone and porcine femora were performed after 1 min and 1h.
RESULTS: During screw insertion, measured peak pressures (>40 MPa) exceeded the compressive failure stress of metaphyseal bone by more than an order of magnitude. Using a bone wedge between tendon and screw reduced local peak pressure by 85% but produced also inferior average contact pressure. In all approaches, initially achieved graft contact pressure rapidly decreased to approximately 25% within 30 min. Pullout strength was significantly reduced after 1h in comparison to 1 min in porcine bone as well as Sawbone.
CONCLUSION: Viscoelastic adaptation of the tendon is severe and critically reduces effective graft-bone contact pressure. Consideration of this newly recognized effect may open new and improved approaches for tendon graft fixation.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22197631     DOI: 10.1016/j.knee.2011.11.005

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  3 in total

1.  Interference screws should be shorter than the hamstring tendon graft in the bone tunnel for best fixation.

Authors:  Michael Stalder; Mazda Farshad; Jess G Snedeker; Dominik C Meyer
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-03-21       Impact factor: 4.342

2.  Interference screw fixation of free tendon grafts: significant time-dependent decrease of the initial contact forces.

Authors:  Christian Stärke; Ulrike Winkelmann; Siegmar Glüge; Roland Becker; Christoph H Lohmann; Uwe Winkelmann
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-11-28       Impact factor: 4.342

3.  Femoral fixation methods for hamstring graft in anterior cruciate ligament reconstruction: A network meta-analysis of controlled clinical trials.

Authors:  Shixin Nie; Shuqing Zhou; Wei Huang
Journal:  PLoS One       Date:  2022-09-22       Impact factor: 3.752

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.