E Cendejas-Bueno1, A Forastiero, J L Rodriguez-Tudela, M Cuenca-Estrella, A Gomez-Lopez.
Abstract
In this paper we report on the development and validation of two different methods for posaconazole quantification from serum samples, HPLC/UV and bioassay. Both methods have been validated according to international guidelines and were also applied to the analysis of 61 trough serum samples from treated patients. A good correlation between both methods was observed. The HPLC method, more laborious and expensive, was demonstrated to be more accurate, precise and faster (analytical range 0.125-16 μg/mL, accuracy between -2.48 and 3.70% and precision between 2.77 and 5.93%, with an analytical run time of 11 min), making it a valuable tool for reference laboratories that centralize high numbers of samples. The microbiological method, however, is simple and offers sufficient precision and accuracy (analytical range 0.125-16 μg/mL, accuracy between -8.10 and 3.77% and a precision between 4.52 and 10.07%), to be used to monitor posaconazole. It may be a valid alternative to chromatographic methods in clinical laboratories without specialized facilities.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22192527 DOI: 10.1111/j.1469-0691.2011.03732.x
Source DB: PubMed Journal: Clin Microbiol Infect ISSN: 1198-743X Impact factor: 8.067