Literature DB >> 22190841

Enhancement of mobile C-arm cone-beam reconstruction using prior anatomical models.

Ofri Sadowsky1, Junghoon Lee, Edward Grant Sutter, Simon J Wall, Jerry L Prince, Russell H Taylor.   

Abstract

We demonstrate an improvement to cone-beam tomographic imaging by using a prior anatomical model. A protocol for scanning and reconstruction has been designed and implemented for a conventional mobile C-arm: a 9 inch image-intensifier OEC-9600. Due to the narrow field of view (FOV), the reconstructed image contains strong truncation artifacts. We propose to improve the reconstructed images by fusing the observed x-ray data with computed projections of a prior 3D anatomical model, derived from a subject-specific CT or from a statistical database (atlas), and co-registered (3D/2D) to the x-rays.The prior model contains a description of geometry and radiodensity as a tetrahedral mesh shape and density polynomials, respectively. A CT-based model can be created by segmentation, meshing and polynomial fitting of the object's CT study. The statistical atlas is created through principal component analysis (PCA) of a collection of mesh instances deformably-registered (3D/3D) to patient datasets.The 3D/2D registration method optimizes a pixel-based similarity score (mutual information) between the observed x-rays and the prior. The transformation involves translation, rotation and shape deformation based on the atlas. After registration, the image intensities of observed and prior projections are matched and adjusted, and the two information sources are blended as inputs to a reconstruction algorithm.We demonstrate recostruction results of three cadaveric specimens, and the effect of fusing prior data to compensate for truncation. Further uses of hybrid reconstruction, such as compensation for the scan's limited arc length, are suggested for future research.

Entities:  

Year:  2009        PMID: 22190841      PMCID: PMC3242453          DOI: 10.1117/12.813405

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  5 in total

1.  Image reconstruction from fan-beam projections on less than a short scan.

Authors:  Frédéric Noo; Michel Defrise; Rolf Clackdoyle; Hiroyuki Kudo
Journal:  Phys Med Biol       Date:  2002-07-21       Impact factor: 3.609

2.  Projected tetrahedra revisited: a barycentric formulation applied to digital radiograph reconstruction using higher-order attenuation functions.

Authors:  Ofri Sadowsky; Jonathan D Cohen; Russell H Taylor
Journal:  IEEE Trans Vis Comput Graph       Date:  2006 Jul-Aug       Impact factor: 4.579

3.  Deformable 2D-3D registration of the pelvis with a limited field of view, using shape statistics.

Authors:  Ofri Sadowsky; Gouthami Chintalapani; Russell H Taylor
Journal:  Med Image Comput Comput Assist Interv       Date:  2007

4.  Multi-modal volume registration by maximization of mutual information.

Authors:  W M Wells; P Viola; H Atsumi; S Nakajima; R Kikinis
Journal:  Med Image Anal       Date:  1996-03       Impact factor: 8.545

5.  Statistical atlases of bone anatomy: construction, iterative improvement and validation.

Authors:  Gouthami Chintalapani; Lotta M Ellingsen; Ofri Sadowsky; Jerry L Prince; Russell H Taylor
Journal:  Med Image Comput Comput Assist Interv       Date:  2007
  5 in total
  3 in total

1.  Volume-of-change cone-beam CT for image-guided surgery.

Authors:  Junghoon Lee; J Webster Stayman; Yoshito Otake; Sebastian Schafer; Wojciech Zbijewski; A Jay Khanna; Jerry L Prince; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2012-07-17       Impact factor: 3.609

2.  Behavior of tip-steerable needles in ex vivo and in vivo tissue.

Authors:  Ann Majewicz; Steven P Marra; Mark G van Vledder; MingDe Lin; Michael A Choti; Danny Y Song; Allison M Okamura
Journal:  IEEE Trans Biomed Eng       Date:  2012-06-13       Impact factor: 4.538

3.  Evaluation of Robotic Needle Steering in ex vivo Tissue.

Authors:  Ann Majewicz; Thomas R Wedlick; Kyle B Reed; Allison M Okamura
Journal:  IEEE Int Conf Robot Autom       Date:  2010-05-03
  3 in total