Effect of image value-to-density table (IVDT) on the accuracy of delivery quality assurance (DQA) process in helical tomotherapy.

Ju-Young Song1, Sung-Ja Ahn.   

Abstract

The effect of the accuracy in the application of the image value-to-density table (IVDT) from kilovoltage computed tomography (kVCT) and mega-voltage computed tomography (MVCT) images on the delivery quality assurance (DQA) in helical tomotherapy (HT) was analyzed. The DQA plans were prepared with a kVCT image of a cheese phantom for 10 HT patients, and the difference in absolute dose equivalence between the planned and real measurement was evaluated according to the accuracy of IVDT application. The difference between the calculated dose distribution and real dose distribution measured with MapCHECK (SunNuclear, Melbourne, FL) was analyzed through the DQA process with a kVCT MapCHECK image and the same analysis was performed with an MVCT MapCHECK image. The IVDT for kVCT was applied to MVCT and the variation in error between the planned and real measurement caused by improper application of IVDT was evaluated. The accuracy of the IVDT application in the homogeneous water-equivalent cheese phantom had only a minor influence on the dose calculation. Although the overall accuracy of the calculated dose was increased when the proper IVDT was applied, this result had no statistical significance. The MVCT image of MapCHECK contained less error between the calculated dose and delivered dose with a high pass rate. The proper IVDT application to the MVCT image of MapCHECK increased the accuracy of dose calculation, and this result had a statistical significance. Application of the correct IVDT is important in HT DQA and its significance is increased when using phantoms consisting of inhomogeneous density materials.
Copyright © 2012 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 22189030     DOI: 10.1016/j.meddos.2011.09.006

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  2 in total

1.  Evaluation of the Flash effect in breast irradiation using TomoDirect: an investigational study.

Authors:  Dae Gyu Kang; Sung Ill Park; Sung Hwan Kim; Mi Joo Chung; Kwang-Man Lee; Jong Hoon Lee
Journal:  J Radiat Res       Date:  2015-02-10       Impact factor: 2.724

2.  Teflon cylindrical phantom for delivery quality assurance of stereotactic body radiotherapy (SBRT).

Authors:  Danielle W Lack; Ali Kakakhel; Ross Starin; Michael Snyder
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.