| Literature DB >> 22186949 |
Kevan J T Anderson1, Greig C Scott, Graham A Wright.
Abstract
The purpose of this study is to describe a new active technique for accurately determining both the position and orientation of the tip of a catheter during magnetic resonance (MR)-guided percutaneous cardiovascular procedures. The technique utilizes phase information introduced into the MR signal from a small receive coil located on the distal tip of the catheter. Phase patterns around a small receive coil are rich in information that is directly related to position and orientation. This information can be collected over a large spherical volume with a diameter several times that of the receive coil. The high degree of redundancy yields the potential for an accurate and robust method of catheter tracking. A tracking algorithm is presented that performs catheter tip localization using phase images acquired in two orthogonal planes without any a priori knowledge of catheter position. Associated experimentation demonstrating feasibility is also presented.Mesh:
Year: 2011 PMID: 22186949 DOI: 10.1109/TMI.2011.2179944
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048